Usage | Sample Collection, Preparation, and Storage 1. Serum: After placing whole blood samples at room temperature for 2 hours or at 4°C overnight, centrifuge at 1000×g for 20 minutes. Remove the supernatant for testing. Blood collection tubes should be disposable, pyrogen-free, and endotoxin-free. Store at -20°C or -80°C and avoid repeated freezing and thawing. 2. Plasma: Within 30 minutes of collection, centrifuge at 1000×g for 15 minutes at 2-8°C. Remove the supernatant for testing. EDTA-Na2 is recommended as an anticoagulant. Avoid using samples with hemolysis or hyperlipidemia. Store at -20°C or -80°C and avoid repeated freezing and thawing. 3. Tissue Homogenization: Take an appropriate amount of tissue and wash it in pre-chilled PBS (0.01M, pH 7.0-7.2) to remove blood (lysed red blood cells in the homogenate will affect the measurement results). After weighing, mince the tissue and mix it with the appropriate volume of PBS (generally a 1:9 weight-to-volume ratio; the specific volume can be adjusted according to experimental needs and recorded. It is recommended to add protease inhibitors to the PBS). Pour the mixture into a glass homogenizer and grind thoroughly on ice. To further lyse tissue cells, the homogenate can be sonicated or freeze-thawed repeatedly (keep the sonication in an ice bath and repeat the freeze-thaw cycle twice). Finally, centrifuge the homogenate at 5000 × g for 5-10 minutes. Remove the supernatant for analysis. (The tissue homogenate should also be assayed for protein concentration to obtain a more accurate concentration of the test substance per milligram of protein.) 4. Cell Culture Supernatant: Centrifuge the cell supernatant at 1000 × g for 20 minutes to remove impurities and cell debris. Remove the supernatant for testing and store at -20°C or -80°C, but avoid repeated freezing and thawing. 5. Urine: Collect the first morning urine (midstream) or 24-hour urine, centrifuge at 2000×g for 15 minutes, collect the supernatant, and store the sample at -20°C. Avoid repeated freezing and thawing. 6. Saliva: Collect the sample using a saliva sample collection tube, then centrifuge at 1000×g for 15 minutes at 2-8°C. Remove the supernatant for testing, or aliquot and store at -20°C. Avoid repeated freezing and thawing. 7. Other biological samples: Centrifuge at 1000×g for 20 minutes, remove the supernatant, and store at -20°C. Avoid repeated freezing and thawing.
Precautions 1. The sample should be clear and transparent, and suspended matter should be removed by centrifugation. Hemolysis of the sample will affect the results, so hemolyzed samples should not be used. 2. If the sample is to be tested within one week of collection, it can be stored at 4°C. If testing cannot be done promptly, aliquot the sample into a single-use amount and freeze at -20°C (for testing within one month) or -80°C (for testing within three to six months). Avoid repeated freeze-thaw cycles. Bring the sample to room temperature before the experiment. Sample Dilution Guidelines If your test sample requires dilution, the general dilution guidelines are as follows: 1. 50-fold dilution: One-step dilution. Dispense 5 μL of sample into 245 μL of Standard and Sample Diluent for a 50-fold dilution. 2. 100-fold dilution: One-step dilution. 3. 1000-fold dilution: Two-step dilution. Add 5 μL of sample to 95 μL of standard and sample diluent for a 20-fold dilution. Then, add 5 μL of the 20-fold diluted sample to 245 μL of standard and sample diluent for a 50-fold dilution, for a total of 1000-fold dilution. 4. 100,000-fold dilution: Three-step dilution. Add 5 μL of sample to 195 μL of Standard & Sample Diluent for a 40-fold dilution. Then, add 5 μL of the 40-fold diluted sample to 245 μL of Standard & Sample Diluent for a 50-fold dilution. Finally, add 5 μL of the 2,000-fold diluted sample to 245 μL of Standard & Sample Diluent for a 50-fold dilution, for a total dilution of 100,000-fold. 5. For each dilution step, use at least 3 μL of liquid, and the dilution factor should not exceed 100. Excessively small sample volumes can easily lead to greater errors during mixing. Ensure thorough mixing at each dilution step to avoid foaming. 6. If the dilution factor is very high, you can dilute it with PBS first, and then use the standard and sample diluent provided in the kit as the final step. Sample Dilution Recommendations1. Normal, fresh serum/plasma samples are recommended for testing (Original solution). 2. Due to individual differences, the recommended dilution factor is for reference only. For actual testing, please estimate the sample concentration range in advance and determine the dilution factor of the sample to be tested through preliminary experiments. Preparation for Testing 1. Remove the kit from the refrigerator 30 minutes in advance and equilibrate to room temperature. 2. Dilute the 25× concentrated wash buffer to 1× working solution with double-distilled water. Return the unused portion to 4°C. 3. Standards: Add 1.0 mL of Universal Standard & Sample Diluent to the lyophilized standard. Tighten the cap and let stand for 10 minutes to fully dissolve. Then gently mix (concentration: 100 ng/mL). Then, serially dilute the standard to 100 ng/mL, 50 ng/mL, 25 ng/mL, 12.5 ng/mL, 6.25 ng/mL, 3.13 ng/mL, and 1.57 ng/mL. Use a blank well with the standard diluent (0 ng/mL). Prepare the required amount of standard and set aside. It is recommended that the prepared standard be added to the sample within 15 minutes; it is not recommended to leave it for extended periods. 4. Biotinylated Antibody Working Solution: Before the experiment, calculate the required volume for the experiment (based on 100 μL/well; add 100-200 μL more). 15 minutes before use, dilute the concentrated biotinylated antibody (1:100) with Biotinylated Antibody Diluent to the working concentration for use that day. For dilution, add 1 μL of concentrated biotinylated antibody to 99 μL of Biotinylated Antibody Diluent and mix thoroughly using a pipette. 5. Enzyme Conjugate Working Solution: Before use, calculate the required volume for each experiment (assuming 100 μL/well; add 100-200 μL more). 15 minutes before use, dilute the concentrated HRP enzyme conjugate (1:100) with enzyme conjugate diluent to the working concentration for use that day. To achieve this dilution, add 1 μL of concentrated enzyme conjugate to 99 μL of enzyme conjugate diluent and mix thoroughly with a pipette. 6. TMB Substrate - Pipette the required volume of solution; do not return any remaining solution to the reagent bottle. Precautions: 1. Ensure all components of the kit are dissolved and mixed thoroughly before use. Discard any remaining standard after reconstitution. 2. Concentrated biotinylated antibody and enzyme conjugate solutions are relatively small and may disperse throughout the tube during transportation. Before use, centrifuge at 1000 × g for 1 minute to allow any liquid on the tube walls or cap to settle to the bottom. Mix the solutions by carefully pipetting 4-5 times before use. Prepare the standard, biotinylated antibody working solution, and enzyme conjugate working solution according to the required volume and use the corresponding diluents. Do not mix them. 3. Concentrated wash buffer may crystallize after removal from the refrigerator. This is normal. Dissolve the crystals completely in a water bath or incubator before preparing the wash buffer (do not heat above 40°C). The wash buffer should be at room temperature before use. 4. Samples should be added quickly, preferably within 10 minutes for each addition. To ensure accuracy, replicate wells are recommended. When pipetting reagents, maintain a consistent order of addition from well to well. This will ensure consistent incubation times for all wells. 5. During the wash process, any remaining wash solution in the reaction wells should be patted dry on absorbent paper. Do not place filter paper directly into the reaction wells to absorb water. Before reading, be sure to remove any remaining liquid and fingerprints from the bottom of the wells to avoid affecting the microplate reader reading. 6. The color developer TMB should be protected from direct sunlight during storage and use. After adding the substrate, carefully observe the color change in the reaction wells. If a gradient is already evident, terminate the reaction early to avoid excessive color change that could affect the microplate reader reading. 7. All test tubes and reagents used in the experiment are disposable. Reuse is strictly prohibited, as this will affect the experimental results. 8. Please wear a lab coat and latex gloves for proper protection during the experiment, especially when testing blood or other body fluid samples. Please follow the National Biological Laboratory Safety Regulations. 9. Components from different batches of the kit should not be mixed (except for the wash solution and the reaction stop solution). 10. The enzyme labeling strips in the kit are removable. Please use them in batches according to experimental needs. Procedure 1. Before beginning the experiment, all reagents should be equilibrated to room temperature. Prepare all reagents in advance. When diluting reagents or samples, mix thoroughly, avoiding foaming as much as possible. If the sample concentration is too high, dilute with sample diluent to bring the sample within the detection range of the kit. 2. Add 100 μL of the standard or sample to be tested (if the sample requires dilution, refer to the sample dilution guidelines for dilution methods). Be careful not to create bubbles. Add the sample to the bottom of the ELISA plate well, avoiding contact with the well walls. Gently shake to mix. Cover the plate or apply film, and incubate at 37°C for 80 minutes. To ensure the validity of the experimental results, use a fresh standard solution for each experiment. 3. Discard the liquid in the wells, spin dry, and wash the plate three times. Wash each well with 200 μL of wash buffer, soak for 1-2 minutes, and discard the liquid from the plate (or wash the plate using a plate washer). After the final wash, pat the plate dry on absorbent paper. 4. Add 100 μL of biotin antibody working solution to each well (can be prepared 15 minutes in advance), cover the plate with film, and incubate at 37°C for 50 minutes. 5. Discard the liquid from the wells and wash the plate three times. Wash each well with 200 μL of wash buffer, soak for 1-2 minutes, and discard the liquid from the plate (or wash the plate using a plate washer). After the final wash, pat the plate dry on absorbent paper. 6. Add 100 μL of enzyme conjugate working solution to each well (can be prepared 15 minutes in advance) and incubate at 37°C for 50 minutes. 7. Discard the liquid in the wells and wash the plate 5 times. Wash each well with 200 μL of washing solution, soak for 1-2 minutes, and shake off the liquid in the ELISA plate (or wash the plate using a plate washer). After the final wash, pat the plate dry on absorbent paper. 8. Add 90 μL of TMB chromogenic substrate solution to each well and incubate at 37°C in the dark for 20 minutes (shorten or extend the time as appropriate based on the actual color development, but not exceed 30 minutes). 9. Add 50 μL of stop solution to each well to terminate the reaction (the blue color will immediately turn yellow). The order of adding the stop solution should be as similar as possible to the order of adding the color developer. To ensure the accuracy of the experimental results, the stop solution should be added as soon as possible after the substrate reaction time expires. 10. Immediately measure the optical density (OD) of each well using a microplate reader at a wavelength of 450 nm. The instrument should be preheated and the assay program set before use. Calculation of Results 1. Subtract the OD value of the blank well from the OD value of each standard and sample. If replicate wells are used, the average value should be used for calculation. 2. For ease of calculation, although concentration is the independent variable and OD value is the dependent variable, we use the OD value of the standard as the horizontal axis (X-axis) and the concentration of the standard as the vertical axis (Y-axis) when plotting. To ensure intuitive visualization of the results, the graphs present raw data rather than logarithmic values. Due to differences in experimental operating conditions (such as operator, pipetting technique, plate washing technique, and temperature), the OD values of the standard curve will vary. The provided standard curve is for reference only; experimenters should establish their own standard curve. The sample concentration can be calculated from the OD value of the sample used on the standard curve. This value is then multiplied by the dilution factor to determine the actual sample concentration. It is recommended to use professional curve drawing software, such as curve expert.
Sample Type Recovery Range Average Recovery Serum (n=5) 80-98% 89% EDTA Plasma (n=5) 82-94% 88% heparin plasma (n=5) 84-98% 91% |
The samples spiked with human CEACAM1 were diluted 2-fold, 4-fold, 8-fold, and 16-fold for recovery experiments, and the recovery rate range was obtained
Sample type | 1:2 | 1:4 | 1:8 | 1:16 |
79-93% | 82-90% | |||
EDTA Plasma(n=5) | 95-103% | 89-101% | 87-98% | 85-92% |
heparin plasma (n=5) | 88-102% | 91-99% | 83-95% | 89-97% |
Chinese Name | 96T | Storage Conditions |
ELISA Plate (Detachable) | 12 strips x 8 wells | |
Lyophilized Standards | 2 | 4°C/-20°C |
Standards & Sample Dilutions | 20 mL | 4°C/-20°C |
Concentrated Biotinylated Antibody (100×) | 120 μL | 4°C/-20°C |
Biotinylated Antibody Dilution Buffer | 12 mL | 4°C/-20°C |
Concentrated HRP enzyme conjugate (100×) | 120 μL | 4°C/-20°C |
Enzyme Conjugate Dilution Buffer | 12 mL | 4°C/-20°C |
Concentrated Wash Buffer (25×) | 20 mL | 4°C/-20°C |
Chromogenic substrate solution (TMB) | 10 mL | 4°C/-20°C (protect from light) |
Reaction stop solution | 6 mL | 4°C/-20°C |
Sealing film | 2 | Normal temperature |
1. If the entire kit is stored at -20°C, please place the kit at 4°C the night before the experiment.
2. Salt precipitation may occur when the concentrated wash solution is stored at low temperatures. When diluting, warm it in a water bath to help dissolve it.
3. A small amount of water-like substance may be present in the wells of a newly opened ELISA plate. This is normal and will not affect the experimental results.
4. This kit is for laboratory research and development use only and is not intended for use on humans or animals.
5. Reagents should be treated as hazardous substances and should be handled with care and disposed of properly.
6. Always wear gloves, lab coats, and protective glasses to avoid contact between skin and eyes with the stop solution and TMB. If contact occurs, rinse thoroughly with water.