The Direct-Pure UP Ultrapure & RO Lab Water System is an integrated system that produces ultrapure and reverse osmosis (RO) water from tap water directly. Quality of ultrapure water meets or exceeds ASTM, CLSI, CAP, and ISO Type I water standards.
Features:
- Electrical and water compartments are completely segregated, thus safer to operate.
- The fully automated control system contains pre-rinse, RO membrane fouling preventing flush, RO membrane and cartridge life monitoring as well as many other programs for convenient operation.
- The pre-rinsing program starts automatically when a new RO membrane is installed so there is no need for users to set up additional cleaning procedures.
- Two sets of dual-column purification cartridges work sequentially to ensure thorough removal of ions and organics in water.
- The system recirculates automatically every hour to prevent microbial from growing
- Resistivity is measured by a high-precision resistivity meter.
- The system is equipped with a 30-liter PE tank (upgradable to a 60-liter) and a liquid level sensor.
- A 0.2 μm final filter or ultrafiltration filter (optional) removes final contaminants.
- RephiLe can provide full document support to meet users’ GMP, GLP, FDA and other certification requirements.
Main Applications:
Ultrapure Water |
RO Pure Water |
HPLC mobile phase preparation
Preparation of reagent blank solution
As sample diluents for GC, HPLC, AA, ICP-MS and other analytical techniques
Preparation of buffer and culture media for mammalian cell culture
Preparation of molecular biology reagents, etc. |
Glassware cleaning
Washing machine for glassware
Water bath water
Autoclave
Feed water to laboratory animals |
Specification:
Feed Water Quality |
|
Tap water* |
Municipal water, TDS < 1000 ppm (2000 μS/cm) |
Operating Temperature |
5 - 35 ℃ |
Feed Water Pressure |
15 - 90 psi (1 - 6 bar) |
Water Production Rate |
|
Ultrapure water (Type I) |
Up to 2.0 L/min at 25℃ |
RO water |
10 to 30 L/h at 25℃ |
Product Water Quality |
|
RO water conductivity** |
Typically < 20 μS/cm |
Resistivity |
18.2 MΩ·cm at 25℃ |
TOC |
< 10 ppb (< 5 ppb with a dual wavelength UV lamp) |
Particle (> 0.2 μm) |
No particles (with a 0.2 μm final filter) |
Microorganisms |
< 0.01 cfu/ml (with a 0.2 μm final filter) |
Pyrogens (endotoxins) |
< 0.001 Eu/ml (with a RephiBio filter) |
RNase |
< 0.5 pg/ml (with a RephiBio filter) |
DNase |
< 10 pg/μL (with a RephiBio filter) |
Dimensions |
|
Length x Depth x Height |
30 cm × 51 cm ×48 cm |
Net Weight (Operating Weight) |
22 - 25 kg (25 - 32 kg) |
*If water hardness is less than 180 ppm, RephiLe suggests using RephiTEKT RW50211KT.
If water hardness is greater than 180 ppm, RephiLe suggests using an additional water softening device to protect the RO membrane in your system.
** Product water quality may vary depending on local feed water conditions.
Ordering information:
Description |
Catalog No. |
Direct-Pure Water System, UP 10 |
RD0P0100E |
Direct-Pure Water System, UP 15 |
RD0P0150E |
Direct-Pure Water System, UP 20 |
RD0P0200E |
Direct-Pure Water System, UP 30 |
RD0P0300E |
Direct-Pure Water System, UP 50 |
RD0P0500E |
Direct-Pure Water System, UP 10 UV |
RD0P010UE |
Direct-Pure Water System, UP 15 UV |
RD0P015UE |
Direct-Pure Water System, UP 20 UV |
RD0P020UE |
Direct-Pure Water System, UP 30 UV |
RD0P030UE |
Direct-Pure Water System, UP 50 UV |
RD0P050UE |
Cartridge Set is required for the water systems to function. Please contact RephiLe for detailed information.
Contact us for more information
US: call toll-free 1-855-RephiLe (1-855-737-4453)
Other Countries: email info@rephile.com. A local representative in your area will contact you.