• Centralized purification and distribution system capable of supplying pure water to one or more laboratories on a loop system
• Features a 200 liters per hour reverse osmosis module, UV photo-oxidation and a feed to and from a deionization cylinder
• Incorporates a wide range of safety features and alarms to ensure pure water quality is never compromised
• Standard CENTRA systems deliver a loop flowrate of up to 18 liters per minute. The CENTRA-R 200 is also available as high flow variable speed model, (HFV) capable of delivering up to 30 liters per minute
• Compatible with building management systems
Standard CENTRA Model |
R 200 |
Loop Flow rate |
Up to 18 liters per minute |
High Flow CENTRA Model |
R 200-HFV |
Typical Loop Flow Rate |
Variable speeds, up to a maximum of 30 liters per minute 1 |
Reservoir Capacity |
350 liters |
Recirculation Pump |
√ |
Reverse Osmosis |
√ |
UV Photo-oxidation |
√ |
0.2μm Filter |
√ |
Optional Deionization |
√ |
Product Water Specifications
The performance specifications given below are intended only as a guide.
Each system can be individually specified to meet your particular requirements. |
Inorganics (resistivity @ 25°C) |
Up to 18.2 MΩ-cm 3 |
TOC – typical |
<10 ppb 3 |
Bacteria – typical 2 |
<5 CFU/ml 2 |
Subject to suitable operation, maintenance and sanitization
1 Subject to the distribution loop design
2 Subject to correctoperating and maintenance procedures
3 Subject to use of suitable deionization cylinder
4 Subject to feedwater
bio-equip.cn