• Designed to directly feed individual or multiple clinical analyzers with up to 4 liters per minute of CLRW grade (CLSI) water
• High flux reverse osmosis with a make-up rate of 30, 60 or 120 liters per hour
• Low bacterial specification – UV and microfiltration technologies together with recirculation and a simple anitization regime keep bacterial contamination below 1 CFU/ml
• Protek high capacity adsorption pretreatment cartridge avoids the need for external filters
• Medpure long life deionization cartridge reduces exchange frequency and running costs
• E-Cartridge on the MEDICA Pro-RE for the reduction of CO2
Applications
• Feed to clinical chemistry analyzers
• Feed to immunodiagnostic analyzers
MEDICA Pro Model |
Pro-30 |
Pro-60 |
Pro-120 |
Inorganics (resistivity @25°C) |
>10 MΩ-cm |
>10 MΩ-cm |
>10 MΩ-cm |
Organics (TOC) – typical |
<30 ppb |
<30 ppb |
<30 ppb |
Bacteria – typical |
<1 CFU/ml |
<1 CFU/ml |
<1 CFU/ml |
Subject to correct operating and maintenance procedures