Unique Human Voice Alarm system
|
The device triggered the human voice prompt and stop injection automatically when there is any malfunction, making the infusion process safer and more reliable.
|
Injection Rate:
|
50/60ml Syringe: 0.1ml/h~999.9ml/h (0.1ml/h step)
1,000ml/h~1,500ml/h (1ml/h step)
30ml Syringe: 0.1ml/h~900.0ml/h (0.1ml/h step)
20ml Syringe: 0.1ml/h~600.0ml/h (0.1ml/h step)
10ml Syringe: 0.1ml/h~300.0ml/h (0.1ml/h step)
|
Flow Rate Accuracy:
|
Within �3% (after correct calibration)
|
Mechanical precision:
|
Within �2%
|
Bolus Rate:
|
50/60ml Syringe: 1,200ml/h
30 ml Syringe: 720ml/h
20 ml Syringe: 480ml/h
10 ml Syringe: 240ml/h
|
Purge Rate:
|
50/60ml Syringe: 1,500ml/h
30ml Syringe: 900ml/h
20ml Syringe: 600ml/h
10ml Syringe: 300ml/h
|
Volume Limit:
|
0.1ml~999.9ml
|
Total Injection
Volume:
|
0.1ml~9999.9ml (0.1ml step)
|
Occlusion:
|
High: 800mmHg �200mmHg (106.7kPa�26.7kPa)
Medium: 50/600mmHg �100mmHg (66.7kPa�13.3kPa)
Low: 300mmHg �100mmHg (40.7kPa�13.3kPa )
|
Alarms:
|
Injection soon finish, end of injection,occlusion, improperly installation of syringe, wrong setting, low battery, syringe loose etc.
|
Power Source:
|
AC 100V~240V, 50/60Hz; Internal rechargeable Li battery, capacity=1,600mAh, 4 hours internal battery backup
|
KVO rate:
|
1ml/h
|
Fuse:
|
F1AL/250/60V, 2pcs inside
|
Power Consumption:
|
30VA
|
IP Classification:
|
IPX4
|
Equipment Classification:
|
Class I, internal power supply, Type CF
|
Operating Condition:
|
Ambient temperature: +5?~+40?;
|
Relative humidity:
|
20~90%
|
Transport & Storage Condition:
|
Ambient temperature: -30?~+55?
|
Relative humidity:
|
=95%
|
Dimension:
|
280mm (L) � 210mm (W) � 130mm (H)
|
Weight:
|
2.2kg (net weight)
|