Name
|
Real-time PCR Detection System
|
1
|
Model
|
AFD4800
|
2
|
Sample Size
|
48×0.2ml centrifugal tube(10-100μl), adapted to 6×8 tube in connection
|
3
|
Reaction System
|
10~1010Copies
|
4
|
Fluorescent Dye
|
F1:FAM、SYBR GreenⅠand others
|
5
|
F2:VIC、HEX、JOE and others
|
6
|
Fluorescence Detection Time of Full Plane
|
<3s(F1+F2)
|
7
|
Temperature Range
|
30.0-99.9℃
|
8
|
Uniformity
|
±0.3℃(@55℃)
|
9
|
Temperature Accuracy
|
±0.1℃(@55℃)
|
12
|
Heating Rate (Mean)
|
≥1.8℃/s
|
13
|
Cooling Rate (Mean)
|
≥1.8℃/s
|
14
|
Temperature Range of Hot-Lid
|
30-105℃
|
15
|
Fluorescence Intensity Detection Repeatability
|
CV≤3%
|
16
|
Sample Detection Repeatability
|
CV≤3%
|
17
|
Sample Linear
|
Linear regression coefficient r≥0.980
|
18
|
Fluorescence Linear
|
Linear regression coefficient r≥0.990
|