Market Positioning
Professinal, Expert, especially designed for clinical diagnosis in hospital.
Design Concept
Accurate, reliable and practical.
Features
1.6 regional independent temperature control and commonlyused functions of quantitative analysis and melting analysis.
2.No time difference for 96 holes florescence detection, achieving high speed, high sensitivity and multichannel detection
3.Advanced fiber optics and imaging analysis with multiple calibration function.
4.Controlled via USB interface
5.OFFICE 2007 interface style, consistent with user habit.
6.Own database and LIS system interface function
7.Custom report printing function
Parameter
1
|
Name
|
Real-time Fluorescence Detection System for Nucleic Acid Amplification
|
2
|
Model
|
AFD9600
|
3
|
Sample Size
|
96×0.2ml Centrifuge tube
|
4
|
Reaction System
|
10-1010Copies/ml
|
5
|
Excitation Light Wavelength
|
450-800nm
|
6
|
Detection Light Wavelength
|
500-800nm
|
7
|
Applicable Dyes
|
F1:FAM,SYBR Green I
|
8
|
F2:VIC,HEX,JOE,TET
|
9
|
F3: ROX,Texas Red
|
10
|
F4:Cy5
|
11
|
F5:Cy5.5,Quasar705
|
12
|
F6-F8 reserved for customers
|
13
|
Module Temperature Control Range
|
4.0-99.9℃
|
14
|
Module Temperature Uniformity
|
±0.3℃
|
15
|
Temperature Control Accuracy
|
±0.2℃
|
16
|
Heating Rate
|
≥2.5℃/s(从50℃-90℃)
|
17
|
Cooling Rate
|
≥2℃/s(从90℃-50℃)
|
18
|
Working Temperature of Hot-lid
|
104℃
|
19
|
Fluorescence Intensity Detection Repeatability
|
CV≤3%
|
20
|
Sample Detection Repeatability
|
CV≤3%
|
21
|
Sample Linear
|
Linear Regression Coefficient r ≥0.98
|
22
|
Fluorescence Linear
|
Linear Regression Coefficien r ≥0.99
|
23
|
Programming
|
20 program segments can be set. Each program segment can be set up to 20 temperature control program section, and maximum cycle is 99.
|
23
|
Communication Interface of Instrument
|
USB 2.0
|
25
|
Operation System
|
Windows XP SP3/Vista /7/8
|
26
|
Input Power
|
100-240V~ 50-60Hz 1000VA。
|
27
|
Size
|
558 × 398 × 300(mm,L×W×H)
|
28
|
Net Weight(Kg)
|
25(excluding computer)
|