Contamination-free two-channel pumping in a compact format—ideal for analyzer, flow cell, and perfusion applications |
 |
Flow range: 0.002 to 12.3 mL/min per channel depending on drive rpm and tubing size |
 |
One- or two-channel, reversible pumping |
 |
Stackable, compact 1/4-DIN housing—mount in your equipment rack or place on your lab bench. (To mount, order brackets 77120-03 at right). |
 |
Separate single-turn speed control and cw/off/ccw switch with green LED power indicator—maintain speed setting when you turn pump on/off or reverse direction. |
 |
Reversible permanent magnet DC motor; purge fluid before or after pumping; pump fluid in either direction. |
 |
Remote control connections on back of pump; start/stop pump with a contact closure. |
 |
揗ax? button runs pump at maximum rpm—prime or rapidly flush tubing. |
 |
Pump head is constructed from polyphenylene sulfide (PPS), stainless steel (SS), and features acetal co-polymer rollers. PPS and SS promote longer life and the six roller construction provides a low-pulsation flow. |
 |
Pump head has average fixed occlusion for tubing; no need to adjust occlusion when changing tubing. |
C/L pumps are compact (1⁄4 DIN) and perfect for low-flow OEM use or analyzer applications. These easy-to-use, economical pumps feature reversible pumping, remote control capabilities, and rapid tubing changes. Ideal for consistent fluid delivery in chromatography, spectroscopy, and dispensing applications.
Maximum suction lift of 5 ft (1.5 m); max discharge head of 20 ft (6.1 m); max pressure 15 psi (1.0 bar). Selfpriming pumps withstand minor back pressure. |
What's included: 5 ft (1.5 m) of 0.89-mm ID Tygon LFL tubing and dual-voltage (115/230 VAC) power supply with 6-ft (1.8-m) power cord. Pumps are shipped with cord/plug assembly appropriate to destination country. Power supplies are UL, CSA, and GS approved. |