Water-jacketed temperature control guarantees precise temperature regulation over the range(IW series)
Speed adjustable fan and the aerating device make it ideal for incubating at temperature slight higher than ambient(IN series)
Speed adjustable fan, aerating device and the unique design of air circulation provides uniform air movement, improved temperature distribution, and makes it ideal for incubating at temperature slight higher than ambient(IN series).
Water-jacketed temperature control guarantees precise temperature regulation over the range, improved temperature distribution within the chamber, and maintains the temperature un-disturbed for several hours after power failure(IW series).
Full features to support safe and convenient operation: water level indication, draining valve, over-flow connection and run-dry protection device for the heater(IW series).
Microprocessor controlled maximum, minimum temperature and opened-door cut-off and alarm functions protect the instrument and your samples from being over-heated, and auto-restart capability after power failure keeps your test un-interrupted.
Hair-style polished stainless steel interior and rounded corner bottom make the daily clean easy for corrosive-resistant long life operation. Adjustable shelves help to make more space for different height.
Specification:
Models
|
XT5106-IN50 XT5106-DHP9052
|
XT5106-IN80
XT5106-DHP9082
|
XT5106-IN150
XT5106-DHP9152
|
XT5106-IN160
XT5106-DHP9162
|
XT5106-IN250
XT5106-DHP9252
|
XT5106-IN270
XT5106-DHP9272
|
Effective capacity
|
L
|
50
|
80
|
150
|
160
|
250
|
270
|
Interior size
|
mm
|
350×350
×410
|
400×400
×500
|
515×450
×640
|
500×500
×650
|
515×450
×1060
|
600×600
×750
|
Overall dimensions
|
mm
|
460×495
×630
|
510×545
×750
|
625×596
×900
|
610×645
×900
|
625×596
×1320
|
710×745
×1000
|
Heater wattages
|
W
|
300
|
300
|
500
|
500
|
800
|
800
|
IN50/IN80/IN150/IN160/IN250/IN270 Equivalent to DHP9052/DHP9082/DHP9152/DHP9162/DHP9252/DHP9272
Dimensions are expressed in width×depth×height.
Models
|
XT5106-IW50 XT5106-GHP9050
|
XT5106-IW80
XT5106-GHP9080
|
XT5106-IW150
XT5106-GHP9150
|
XT5106-IW160
XT5106-GHP9160
|
XT5106-IW250
XT5106-GHP9250
|
XT5106-IW270
XT5106-GHP9270
|
Effective capacity
|
L
|
50
|
80
|
150
|
160
|
250
|
270
|
Interior size
|
mm
|
350×350
×410
|
400×400
×500
|
515×450
×640
|
500×500
×650
|
515×450
×1060
|
600×600
×750
|
Overall dimensions
|
mm
|
460×495
×630
|
510×545
×750
|
625×596
×900
|
610×645
×900
|
625×596
×1320
|
710×745
×1000
|
Heater wattages
|
W
|
800
|
1,000
|
1,600
|
1,600
|
2,000
|
2,000
|
IW50/IW80/IW150/IW160/IW250/IW270 Equivalent to GHP9050/GHP9080/GHP9150/GHP9160/GHP9250/GHP9270
Dimensions are expressed in width×depth×height.
Temperature range
|
℃
|
RT+5~+65
|
Temp stability
|
℃
|
±0.2~0.5
|
Display resolution
|
℃
|
0.1
|
Digital setting
|
℃
|
0.1
|
Shelves (stand/max)
|
pcs
|
1/2
|
2/3
|
2/3
|
2/3
|
3/4
|
3/4
|
Interior materials
|
SUS430 hair-style polished stainless steel
|
Housing materials
|
SPCC steel with powder coating
|
Front door
|
PU foaming insulated door with glass window
|
Inner door
|
Tempered glass, option for front door with no window
|
Duty cycle
|
Continuous
|
Auxiliaries
|
Temperature sensor failure cutoff and alarm, high-temperature cutoff and low temperature alarm, auto-restart after power failure, ATC capability and water heater run-dry protection
|
Optional devices
|
RS232/485 communication interface, side cable hole, power socket inside the incubator and overhead radiating light
|
Power
|
AC220V±10%, 50Hz
|
bio-equip.cn