1. Features of x-ray diffraction analyzer (XRD)
● High stable X-ray source: The instrument uses an X-ray generator to improve the stability of the diffractometer measurement results;
● Long life X-ray tube: metal ceramic X-ray tube, with good heat dissipation, high operating power (40kV×40mA), and long service life characteristics;
● High precision goniometer: The diffraction angle drive, adopting stepper motor drive plus with optical coding control technology, has accurate positioning and measurement; The hidden design of the goniometer makes the instrument look more neat, beautiful, and generous; In the measuring range of diffraction angle, the linearity is less than 0.02°.
● The accessories to broaden the application field of the diffractometer are complete: the installation of high-temperature, low-temperature, multi-functional accessories achieve "plug and play". The software automatic identification technology is convenient for operators to use the instrument;
● Safety and reliability of X-ray protection: The X-ray protection device is safer and reliable, the X-ray protection door is forbidden to open when the sample is measured, and the double protection can protect the operator from radiation.
- Rated output power: 2KW (standard)
- Voltage: 10 ~ 50kV
- Current: 5 ~ 40m
- Stability of voltage: better than ±0.005%
(when outer voltage fluctuation is 10%)
- X-ray tube: metal ceramic X-ray tube
- Focus size: 1×10mm
- Cu target: power 2.4kW (standard configuration).
- High precision goniometer
- Goniometer: θ/θ sample is horizontally placed,
- Adopts incremental encoder reading technology
- Diffraction radius: 225 (standard configuration)
- Scanning mode: continuous, step-by-step
- 2θ Measuring range: -10° ~ 160° (standard configuration)
- Minimum step-angle: 0.0001°
- Maximum corner speed 20°/s,
- Measurement linearity of diffraction angle ≤0.01°
- Repeatability error ±0.0001°
- Angle reproducibility: 0.0001°
- Maximum positioning speed: ≤1200°/min
- Measurement linearity of diffraction angle ≤0.01°
- Scintillation detector: (standard configuration)
- It is necessary to configure a monochromator sometimes to effectively remove the Kβ line,
- Maximum linear count: ≥5×105CPS,
- Spectral resolution: ≤50%.