Technical Feature
●The host adopts a 7-inch color touch screen, with a user-friendly human-machine exchange interface.
● Computer reverse control and host touch screen achieve synchronous bidirectional control.
●A multi-core, 32-bit embedded hardware system ensures the reliable operation of the instrument.
● Digital display of carrier gas/hydrogen/air path flow (pressure).
●Equipped with air shortage alarm protection function; Heating control protection function (when opening the column box door during use, the column box fan motor and heating system automatically turn off).
●The split flow rate/split ratio can be automatically controlled to save carrier gas.
●Configure the installation and positioning interface of the automatic sampler, which can match various specifications of automatic samplers.
●The data collection is equipped with a standard dual channel data collection card, with a sampling time of 50ms.
●Using a logarithmic amplifier board, the detection signal has no cut-off value, and the synchronous external trigger function can be used to start the host and workstation simultaneously by external signals (such as automatic samplers, thermal analyzers, etc.).
●Equipped with comprehensive system self inspection function and automatic fault identification function.
●Equipped with an extended functional interface for 8 external events, it can be equipped with various control valves and operate according to its own set timing.
●The external connection method is network port connection (RJ45), providing convenience for remote control of data.
Column temperature box:
●Temperature range: 5 ℃~400 ℃ above room temperature
Temperature control accuracy: ± 0.1 ℃
● Program heating up: 9th order/10th platform
●Total program time: 9999.9 minutes
●Heating rate (MAX): 60 ℃/min
Sampler:
●Temperature range: 7 ℃~420 ℃ above room temperature
●Can be installed with filling columns and split/non split injectors
●Working in constant voltage mode
●At most three injection systems can be connected simultaneously
Hydrogen flame ionization detector (FID):
●High voltage switch control
●Baseline signal display
●Ignition coil control
●Detection limit: ≤ 3 × 10-12 g/s
●Sample: n-neneneba hexadecane (detection amount: 3pg/s)
●Dynamic range: 107
if need price,please email us to