Features
1.Ideal system for STAT Biochemistry and Electrolyte analysis
2.Fully automated, flexible, reliable and accurate
3.Analysis of electrolytes and chemistries: GLU、ALB、 BUN、Cr 、K+、Na+、CL-、Ca2+ 、CO2, Results obtained within 5 minutes
4.300 tests per hour, STAT sample priority
5.User defined profiles
6.Automatic washing of reaction cuvette
7.Automatic probe cleaning , liquid level detection and collision protection
8.Automatic sample re-run
9.Refrigerated reagent compartment
10.Real-time reagent residual volume checking
11.Real-time monitoring and display of working status and reaction curve
12.Low water consumption
Specifications
Optical system
Light source: Halogen lamp 12V/20W
Interference filter: 340 nm、405 nm、492 nm、510 nm、546 nm、578 nm、630 nm
Half bandwidth: 8nm~12nm
Absorbance range: 0~2.5000 ABS
Absorbance precision: 0.001 ABS
Stray light: < 1.0%(340 nm)
Sample system
Sample tray: 30 positions
Sample volume: 2.0~45.0µL , step by 0.1µL
Sample probe: liquid level detection and collision protection
Probe cleaning: internal and external washing automatically
Reagent system
Reagent tray: 11 positions in refrigerated compartment (2°C ~8°C)
Reagent volume: 20.0~700.0µL , step by 0.5µL
Reagent probe: liquid level detection and collision protection
Probe cleaning: internal and external washing automatically
Reaction system
Reaction cuvette: 28 positions with automatic washing
Reaction volume: 300~1000µL
Flowcell: Stainless steel with quartz window, optical length 10mm , 32µL
Carryover: < 1.0%
Reaction temperature: 37°C
Operation unit
Software system: Windows 2000/ Windows XP
Interface: RS-232
Analysis system
Methodology: ISE End-point, Initial rate, Kinetic
Throughput: 240 tests/hour ,STAT sample priority GLU、ALB、 BUN、Cr 、K+、Na+、CL-、Ca2+ 、CO2 results obtained within 5 minutes
Programming : user defined profiles
Calibration mode: linear (one point and multipoint) / non-linear multipoint
Calibration type: Startup calibration , real-time calibration , repeated calibration
Control: different levels of QC files, L-J graph and data, over-limit alert
Working conditions
Power supply: 220V ± 10%,50/60Hz
Temperature: 15°C~30°C,
Humidity: 20%~80%
Water consumption: 2L / hour
Dimension: 890 mm × 620 mm ×1130 mm (W×D×H)
Weight : 115 kg
bio-equip.cn