Tanon 5200 Chemiluminescent Imaging System
The Tanon – 5200 system is the ideal economical chemiluminescent imager. It uses a CCD camera with high quantum efficiency.The camera is cooled to absolute temperatures to minimize background noise and and enhance the Signal-to-Noise radio. F/0.95 lens upgrade the sensitivity. The system produce chemiluminescent images of faint samples at speeds faster than exposure time of film-based detection.
Features
1. Supercooling to -45℃(absolute) for optimal imaging with low-light applications.
2. True 16-bit date and a dynamic range that covers 4 orders of magnitude to maximize limit of detection.
3. The operation of excitation sources , motorized lens , the imaging taken is controlled by the GIS 1-D software.
Specifications
1. Camera : High-sensitivity monochrome camera
2. Sensor : Sony ICX694
3. Camera cooling system : Peltier
4. Cooling range (absolute) : -45℃
5. Quantum efficiency : High QE: 75%
6. Dark current : <0.0005 e-/pixel/sec. @ -40º C
7. Readout noise : 5.5e- RMS at 12 MHz
8. Exposure time : 1ms-60min
9. Pixel array size (H×V) : 1434 X 1050
10. A/D conver : 16 bit (0-65535)
11. Pixel size (H×V) : 4.54um×4.54um
12. Binning : 1×1,2×2,4×4
13. Dynamic range : ﹥4.6 OD
14. Motorized lens : F/0.8,1 inch lens, Software - controlled Focus
15. Zoom : Two locality sample platform
16. Excitation sources : Overhead LED white light
17. Filter station : No
18. Filter : No
19. Sample area : 16×16cm
20. Operating system compatibility : Windows Xp, Windows 7, Windows 8
bio-equip.cn