Instrument specifications and parameters :
The output flow rate: 0-300 - ml/min;
Output pressure: 0-0.4 Mpa
The pressure stability: < 0.2%
Hydrogen purity: > 99.999%
The maximum power: 150 w
Output interface: 3 mm or 1/8 (M8 in x 1 external thread)
Secondary protection pressure: 0.5 MP
Tank volume: 1.5 liters
Water consumption: 15 ml/h
Quality requirement: resistivity 1 m or more Ω/cm
Power supply voltage: AC 220 v (50/60 Hz) insurance tube: 250 v / 2 a
Working conditions: temperature 10 to 40 ℃, relative humidity 85%, no serious dust
Dimension: 420 * 200 * 200 (L * W * H)
Working pattern: continue to work
Package weight: about 14 kg
The instrument features:
The equipment is mainly used for the air products form a complete set of gas chromatograph, in petroleum, chemical industry, public security, environmental protection, food, commodity inspection, widely used in such fields as science and education, medicine, technology, applies to Agilent Shimadzu, Thermo markets such as the mainstream of imported and domestic gas chromatograph. Hydrogen generator is mainly composed of electrolytic system, pressure control system and purification system. Electrolytic hydrogen is the use of advanced membrane separation technology, the infrared electric device to control the pressure, can make hydrogen production according to the demand of automatic adjustment, stable pressure. Alternative to high pressure hydrogen cylinder, the use of safe and convenient, easy to operate. Generator built-in prevent return liquid device, which can effectively prevent liquid. Produce gas can be continuous or intermittent use, stability, purity is not easy to decay; Over-voltage protection device, safe and reliable.