Instruction |
|
TS-1A/TS-2A are four-channel micro syringe pump. It can hold µL unit standard glass syringe. The features of accurate distance control and broad linear speed range (7.9 µm/min - 79.4 mm/min) can meet versatile requirements. The drive unit is independent. It is easy to combination and installation. Its compact size and vertical (horizontal) installation structure makes this pump easily used in micromanipulator, stereotaxic instrument for various biologic research applications. |
 |
Specifications |
|
- Max. Infusion Distance: 70 mm
- Acceptable Glass Syringe : 5 µL - 1000 µL
- Linear Speed: 7.9 µm/min - 79.4 mm/min
- Adjusting Resolution: 7.9 µm/min
- Distance Resolution: 0.165 µm
- Linear Force: > 20 N
- Setting Mode: Membrane keypad and rotary coded switch
- Accuracy: < +/- 0.5% error in the condition of > 30% of max. infusion distance
- Display: 128 x 64 graphic LCD
- External Control: Start/Stop control, fast forward control, fast reverse control
- Communication Interface: RS 485
- Power: AC 100 V - 240 V or DC 12 V
- Power Consumption: < 40 W
- Operating Condition: Temperature 0 - 40 °C
- Relative Humidity < 80 %
- Controller Dimensions (L x W x H): 170 x 108 x 65 (mm)
- Controller Weight: 0.8 Kg
- Drive Unit Dimensions (L x W x H): 180 x 46 x 78 (mm)
- Drive Unit Weight: 0.6 Kg
- IP Rating: IP 21 |
 |
Functions and Features |
|
- Parameters Setting: The parameters of each channel can be different.
- Running Control: Each drive unit can be controlled separately, or four drive units can be controlled to run simultaneously or run at different time
- Channel Copy: All drive units can be run according to the parameters of one drive units of them
- Delaying Startup: Delaying startup time of each channel can be controlled separately
- Memory Function: Select resume operation or remain stopped when power turns after an interruption
- Block Protection: When one drive unit stops accidently, the system will warm and stop
- Working Mode: TS-1A/L0107-1A - Infusion; TS-2A/L0107-2A - infusion, withdrawal, infusion/withdrawal, withdrawal/infusion, continuours
- External Control: Start/stop input control signal which is pulse mode to switch the states of start and stop
- Communication: Realize computer control through RS485 communication interface
Main Functions for Each Channel
- Syringe Selection: The syringe can be selected in the manufacture table which includes manufacturer, material and size
- User-Defined Glass Syringe: Save 4 inner diameter of user-defined glass syringe barrel
- Parameters Setting: Set dispensing volume, infusion time, pause time and copy number
- Display Mode Selection: Different parameters (column, flow rate, linear speed)can be selected in the main display inter face
- Fast Forward & Fast Reverse: Infusion at the max. speed
- Calibration: Acquire accurate volume through calibration |
 |
Parameters |
|
Syringe Pump |
Glass Syringe
(µL) |
Barrel Inner
Diameter (mm) |
Effective Stroke
(mm) |
Flow Rate (/min) |
Syringe Material |
Weight (kg) |
TS-1A/L0107-1A
or
TS-2A/L0107-2A
|
5 |
0.35 |
51.97 |
0.764 nL - 7.64 µL |
Glass Syringe |
Controller 0.8
Drive Unit
0.6
|
10 |
0.5 |
50.93 |
1.559 nL - 15.59 µL |
25 |
0.8 |
49.74 |
3.989 nL - 39.89 µL |
50 |
1.1 |
52.61 |
7.544 nL - 75.44 µL |
100 |
1.6 |
49.74 |
15.96 nL - 159.6 µL |
250 |
2.3 |
60.17 |
32.98 nL - 329.8 µL |
500 |
3.25 |
60.27 |
65.85 nL - 658.5 µL |
1000 |
4.61 |
59.91 |
132.5 nL - 1325 µL |
|