Air Temperature
|
|
Sensor type
|
Thermistor
|
|
Units
|
°C / °F
|
|
Range
|
0 – +55 °C
|
|
Accuracy
|
±0.5 °C
|
|
Resolution
|
0.1 °C
|
|
Response (90 %)
|
Approx. 7 min
|
|
Relative Humidity
|
|
Sensor type
|
Polymer resistance
|
|
Units
|
%RH
|
|
Range
|
10 – 95 %RH
|
|
Accuracy
|
±5 %RH (at 25 °C, 50 %RH)
|
|
Resolution
|
1 %RH
|
|
Response (90 %)
|
Approx. 7 min
|
|
Internal CO₂ Sensor
|
|
Sensor type
|
NDIR (non-dispersive infrared)
|
|
Units
|
ppm
|
|
Range
|
0 – 9 999 ppm
|
|
Accuracy
|
±(50 ppm + 5 % of reading) ≤ 5 000 ppm
|
|
Resolution
|
1 ppm (min.)
|
|
Response (90 %)
|
Approx. 1 min
|
|
Main Unit
|
|
Memory
|
8 000 data sets (3 ch each)
|
|
Logging interval
|
15 steps: 1 s, 2 s, 5 s, 10 s, 15 s, 20 s, 30 s, 1 min, 2 min, 5 min, 10 min, 15 min, 20 min, 30 min, 60 min
|
|
Logging mode
|
Endless (overwrite) / One-time (stop when full)
|
|
LCD
|
Real-time values (CO₂, T, RH cycled), battery icon
|
|
Interfaces
|
USB (std.), RS-232 (opt.), IR (requires TR-57DCi)
|
|
Transfer time
|
USB ≈ 45 s full memory / IR ≈ 60–80 s
|
|
Alarm output
|
Open-drain MOS: OFF ≤30 V, ON ≤0.1 A, RON ≈15 Ω
|
|
Power
|
AC adaptor (AD-06A1 or AD-06C1) or 4 × AA alkaline (LR6)
|
|
Battery life
|
Approx. 2 days (alkaline only, no adaptor)
|
|
Dimensions
|
96 L × 66 W × 46 H mm (excl. sensor & protrusions)
|
|
Weight
|
214 g (with batteries, excl. sensor)
|
|
Operating environment
|
0 – +45 °C, <90 %RH (no condensation)
|
|
Supplied accessories
|
4 × AA batteries, AC adaptor, USB cable (US-15C), T/RH sensor (THA-3001), software, manual
|