Features
 ※ Large LCD screen(128x64Dots).Can display total 200 groups of data,5 groups per screen. Can display standard curve and the curve equation.
  ※ Can save 200 groups of curve equation. Convenient for check and reload.
  ※ Data can be restored after a sudden power cut.
  ※ Auto setting wavelength. Auto zero and blank.
  ※ Tungsten lamp & deuterium lamp can be turned on/off individually to extend lifetime.
  ※ The Syclon application software UV-Professional provides complete control of the spectrophotometer from a computer through the Built-in USB port. The following functions can be easily achieved: Quantitative, Kinetics, Wavelength Scan, DNA/Protein Test and Multi-wavelength Test.
  ※ The flange of the D2 lamp makes it become true Pre-aligned design makes it convenient to change lamps.
  ※ Large sample compartment, It can accommodate 5-100mm path length cuvettes with  optional holders. A variety of optional accessories are under your choice.
■Specifications
|  Optical System |   Single beam, Grating 1200 lines/mm | 
|  Wavelength Range | 190-1100nm | 
|  Bandwidth | 2nm | 
|  Wavelength Accuracy | ±1nm | 
|  Wavelength Repeatability | 0.5nm | 
|  Wavelength Setting | Auto | 
|  Photometric Accuracy | ±0.3%T | 
|  Photometric Repeatability | 0.2%T | 
|  Photometric Range | -0.3-3A,0-200%T,0-9999C | 
|  Stability | ± 0.002A/h @ 500nm | 
|  Stray Light | ≤0.1%T | 
|  Data Output Port | USB | 
|  Printer Port | Parallel Port | 
|  Display | 128*64 Dots LCD | 
|  Lamps | D2  Lamp  &  W  Lamp | 
|  Detector | Silicon Photodiode | 
|  Power Requirements | AC   220V/50Hz or  110V/60Hz | 
|  Dimension | 460*360*225mm | 
|  Weight | 18kg |