Power Supply
|
100 – 260 VAC, 50HZ, 120W Spare battery(Selection)
|
Operation Pressure
|
100-10000 mbar (145psi) ;
|
Unit
|
mbar
|
Operation Condition
|
Ambient Temperature:+5℃ ~ +40℃;Relative Humidity:10-80%
|
Dimension(mm)
|
455(Depth)*300(Width)*210(Height1)
|
Function
|
Manual Bubble point Test;Basic Bubble Point Test;Extensive Bubble Point Test; Pressure Holding Test;Diffusion FLow Test;Water Immersion Test;Ultrafitration Membrane Test
|
Test Accuracy
|
Upstream Volume Test:≤± 4%;Bubble Point Test:≤± 50mbar ;Diffusion Flow Test:≤± 4% ;Water Immersion:≤± 0.01ml
|
Test Scope
|
BP:100-8000mbar DF:1-600ml/min WI:0.01-100ml/min
|
Applied Range
|
Symmetric and asymmetric membrane test, needle filter, filter bag, flat filter, filter cartridge (big filter volume ), ultrafiltration membrane, ultrafiltration column, a variety of irregular filters
|
Authority Management
|
User Name and Password to Login ,electronic signature to meeting the FDA 21CFR PART 11requirements Audit trail (customization)
|
Reservation Solution
|
120 sets
|
Print Function
|
Built-in printer;
|
Historical Records
|
More than1 5000 sets unlimited sets records
|
Records Copy
|
Support USB disk to export (Include test curve)
|
Display Screen
|
7 inch, color touch screen
|
Serial Connection
|
RS232/USB
|
Language
|
Chinese/English
|
Work Mode
|
Online/Offline
|
Other functions
|
One/two-dimension cods ,Scanning gun;Lithium battery(customization)
|
Applicable Environment
|
Above B level
|
Weight
|
8KG
|