General information |
Machine type |
Random access,open reagent & locked reagent (optional) |
Test speed |
Constant 200 tests/hour (Mono reagent or double reagent) |
Test principle |
Colorimetric method,turbidimetry |
Test method |
1 point end,2 point end,Fixed-Time,Kinetics |
Calibration type |
Linear & nonlinear |
Sample unit |
Sample tray |
40 sample positions |
Sample tube |
Micro cup & test tube & blood collection tube can be used.(φ12 -13 )mm * (25 -100 )mm |
Sample volume |
2-30µL step by 0.1µL |
Sample probe |
Liquid surface detection,timely tracking for liquid volume.Collision avoidance |
Sample barcode |
Sample barcode reader (optional) |
Reagent unit |
Reagent tray |
80 reagent positions (40 for Rl & 40 for R2), Non-stop cooling system with peltier pad inside, 24 hours 2°C-14″C, open system, could accept any reagent |
Reagent volume |
20-300µL,step by 1µL |
Reagent needle |
Liquid surface detection,timely tracking for liquid volume.Collision avoidance. |
Regent barcode |
Regent barcode reader (optional) |
Reaction unit |
Reaction tray |
44 reaction cuvettes, new amorphous optical plastics reaction cups, perfect transmittance. |
Reaction volume |
150µL-330µL |
Reaction time |
10 minutes |
Temperature |
Peltier pad incubation system, reaction temperature (37°C ±0.1 °C) |
Mixing system |
Mixer |
paddle-type mixing,effective reducing of carry-over |
Optical system |
Light source |
Halogen lamp |
Wavelength |
340nm; 405nm; 450nm; 505nm; 546nm; 578nm; 630nm; 700nm(4 more options) |
Absorbance range |
0-4.0 Abs |
Resolution |
0.0001 Abs |
Operating system |
Operation system |
Windows XP, Windows 7, Windows 8, Windows 10 |
Data processing |
Automatic calibration, test combination, effective time management of reagents, serum index, whole process detection, linear expansion of enzyme, blank deduction,dirty cuvette memory , cross infection prevention, patient information memory and association input, automatic report audit, data fuzzy query, report statistics and printing, reference range grading, alarm information classification, user operation privilege classification management |
Report printing |
6 formats optional, support custom pattern |
interface |
TCP/IP network interface |
Working conditions |
Power supply |
AC 110/220V ± 10%,50/60 Hz 650W |
Water consumption |
8L/hour(maximum) |
Dimension |
870mm * 690mm * 740mm(L*W*H) |
Weight |
80kg |