| General information |
| Machine type |
Random access,open reagent & locked reagent (optional) |
| Test speed |
Constant 200 tests/hour (Mono reagent or double reagent) |
| Test principle |
Colorimetric method,turbidimetry |
| Test method |
1 point end,2 point end,Fixed-Time,Kinetics |
| Calibration type |
Linear & nonlinear |
| Sample unit |
| Sample tray |
40 sample positions |
| Sample tube |
Micro cup & test tube & blood collection tube can be used.(φ12 -13 )mm * (25 -100 )mm |
| Sample volume |
2-30µL step by 0.1µL |
| Sample probe |
Liquid surface detection,timely tracking for liquid volume.Collision avoidance |
| Sample barcode |
Sample barcode reader (optional) |
| Reagent unit |
| Reagent tray |
80 reagent positions (40 for Rl & 40 for R2), Non-stop cooling system with peltier pad inside, 24 hours 2°C-14″C, open system, could accept any reagent |
| Reagent volume |
20-300µL,step by 1µL |
| Reagent needle |
Liquid surface detection,timely tracking for liquid volume.Collision avoidance. |
| Regent barcode |
Regent barcode reader (optional) |
| Reaction unit |
| Reaction tray |
44 reaction cuvettes, new amorphous optical plastics reaction cups, perfect transmittance. |
| Reaction volume |
150µL-330µL |
| Reaction time |
10 minutes |
| Temperature |
Peltier pad incubation system, reaction temperature (37°C ±0.1 °C) |
| Mixing system |
| Mixer |
paddle-type mixing,effective reducing of carry-over |
| Optical system |
| Light source |
Halogen lamp |
| Wavelength |
340nm; 405nm; 450nm; 505nm; 546nm; 578nm; 630nm; 700nm(4 more options) |
| Absorbance range |
0-4.0 Abs |
| Resolution |
0.0001 Abs |
| Operating system |
| Operation system |
Windows XP, Windows 7, Windows 8, Windows 10 |
| Data processing |
Automatic calibration, test combination, effective time management of reagents, serum index, whole process detection, linear expansion of enzyme, blank deduction,dirty cuvette memory , cross infection prevention, patient information memory and association input, automatic report audit, data fuzzy query, report statistics and printing, reference range grading, alarm information classification, user operation privilege classification management |
| Report printing |
6 formats optional, support custom pattern |
| interface |
TCP/IP network interface |
| Working conditions |
| Power supply |
AC 110/220V ± 10%,50/60 Hz 650W |
| Water consumption |
8L/hour(maximum) |
| Dimension |
870mm * 690mm * 740mm(L*W*H) |
| Weight |
80kg |