| Product Name | Fluorescence Quantitative PCR Detection System | 
		
			| Model | CYCLER 006 | 
		
			| Sample Capacity | 96-Well PCR plate,12*8-strip,96*0.2ml (Bottom Transparent) | 
		
			| Dynamics Range | 1~1010Copies | 
		
			| Excitation Wavelength | 300-800nm | 
		
			| Emission Wavelength |   500-800mm   | 
		
			| Detected Fluorescence | F1:FAM,SYBR Green I;F2:VIC,HEX,TET,JOE, CY3,NED,TAMRA;F3:ROX,TEXAS-RAD,F4: CY5;F5:CY5.5   | 
		
			| Block Temp.Range |   4-105°C(Minimum Increment:0.1°C) SOAK Low Temp.Conservation Function   | 
		
			| Heating/Cooling rate | 4.0°C/s(max) | 
		
			| Temp.Control Accuracy | ≤±0.1°C | 
		
			| Temp Fluctuation | ≤±0.1°C | 
		
			| Temp Uniformity | ≤±0.3°C | 
		
			| Temp.Control Mode | Block/Tube Simulation Mode (Automatic Control Based On Sample Volume) | 
		
			| Sample Volume Range | 5-100µL | 
		
			| Gradient Temp.Range |   1~36°C   | 
		
			| Hot-lid Temp.Range |   30-110°C(Adjustable,Default 105°C) Automatic Hot-lid   | 
		
			| Fluorescence Detection Repeatability | 5% | 
		
			| Scan Mode |   Entire Plate or Designated Line   | 
		
			| Program |   Max 20 Segments for Each Program,Max 99 Cycles   | 
		
			| Operation Mode | Continue | 
		
			| Scan Period | 5.5s | 
		
			| Feature Function | Absolute Quantification,Relative Quantification,,SNP Analysis;Data Automatic Analysis;Melting Curve Genotyping; Gradient、HRM Mult-channal Crosstalk Correction;Background Correction;Automatic Gain;Customized Parameters   | 
		
			| Operation System |   Microsoft:Windows7/window8.1 Software:Excel2000/2002/2003/2007/2012   | 
		
			| PC Configuration |   Memory:512M Hard Disk:10GB CPU:Pentium®4Virtual Memory:≥1000MB   | 
		
			| Power Supply | 100-240V-50/60HZ 600W | 
		
			| Dimension(L*W*H) |   410mm*386mm*352mm   | 
		
			| Socket | USB Adapter,RS232 Adapter,WIFI Adapter, Bluetooth Adapter | 
		
			|   Authentication |   Ferrotec Pelitier/CE(EMC&LVD)/IVD/RoHS2 /PICC product quality liability insurance |