| neral parameters |
Item |
Specifications |
| Power supply |
AC220V,50/60Hz(Standard) 1320VA; 110V 60Hz(Optional) |
| Size(L*W*H) |
950*645*(1250-11450) mm |
| Maximum heater output power(220VAC) |
Air chamber heater:330W; Water tank heater:155W |
| Auxiliary network power output |
AC220V/50Hz, both maximum permissible currents are 220V 1.5A |
| Heater indication |
0~100%, 10% increasing amount |
| Temperature control mode |
Bed temperature mode; skin temperature mode |
| Bed temperature control range |
25.0~37.0℃; 37.1~39.0℃(Span mode) |
| Skin temperature control range |
34.0~37.0℃;37.1~38.0℃(Span mode) |
| Skin temperature sensor display range |
5.0~65.0℃ |
| Heating time |
≤ 50min |
| Temperature variability |
≤ 0.5℃ |
| Temperature uniformity (mattress in horizontal position) |
≤ 0.8℃ |
| Temperature uniformity (mattress in tilted position) |
≤1.0℃ |
| Difference between average incubator temperature and control temperature |
≤1.5℃ |
| Skin temperature sensor accuracy |
±0.3℃ |
| Humidity control |
The running time of the tank after it is filled with distilled water |
Humidity below 70%RH, at least 24h |
| Water tank capacity |
1500ml |
| Humidity display range |
0%RH~99%RH |
| Humidity control range |
CAH%RH~90%RH(1% per adjustment) |
| Humidity control accuracy |
±10% |
| Oxygen concentration control |
Oxygen concentration display range |
0%~99% |
| Oxygen concentration display resolution |
1% |
| Oxygen concentration control accuracy |
±5% oxygen volume concentration |
| Oxygen concentration setting range |
21% ~ 60% (1% per adjustment) |
| Oxygen sensor service life |
The maximum use time is 10000h at 100% oxygen concentration |
| Transport storage |
Temperature |
-20℃~+55℃ |
| Atmospheric pressure range |
500hPa~1060hPa |
| Relative humidity |
≤93%RH |
| Packing size/ Gross Weight |
Base packing size: 1200*800*870mm 50Kg
Body packing size: 1100*710*855mm 60Kg |
| The work environment |
Relative humidity |
30%RH~75%RH |
| Environment temperature |
+20℃~+30℃ |
| Pressure range |
700hPa~1060hPa |
| Service altitude |
≤2000m |
| Overvoltage category |
Ⅱ |
| Pollution level |
2 |
| Ambient air velocity |
≤ 0.3m/s |
| Other indicators |
Ambient noise in the infant compartment of the incubator |
No more than 55dB(A) at stable temperature |
| Carbon dioxide concentration in the infant compartment |
A mixture of 4% carbon dioxide and air was fed into an 8mm diameter tube in the vertical direction from the mat to the top of the box at 750mL /min to a point 10cm above the mattress center and 10cm above the center. The concentration was less than 0 5% after stabilization |
| The velocity of air flow over the baby mattress |
≤0.35m/s |
General
alarm |
Power failure alarm, fan alarm, sensor alarm, deviation alarm, over temperature alarm, water tank placement error alarm, water shortage alarm |
| System alarm |
Mcu1 and Mcu2 communication error alarm, Mcu2 and Mcu3 communication error alarm, analog-to-digital conversion chip 2544 failure alarm, analog-to-digital conversion chip 2548 failure alarm, over temperature alarm, temperature heating system failure, humidity heating system failure |