| Main Specification | Wavelength range | 190-900nm | 
		
			| Wavelength accuracy | Better than ±0.25nm | 
		
			| Resolution | Two spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley-peak energy ratio less than 30%. | 
		
			| Baseline stability | 0.004A/30min | 
		
			| Background correction | The D2 lamp background correction capability at 1A is better than 30 times.   The S-H background correction capability at 1.8A is better than 30 times. | 
		
			| Light Source System | Lamp turret | Motorized 6-lamp turret (Two high performance HCLs can be mounted on the turret to increase the sensitivity in flame analysis.) | 
		
			| Lamp current adjustment | Wide pulse current: 0~25mA,              Narrow pulse current: 0~10mA.                  | 
		
			| Lamp power supply mode | 400Hz square wave pulse; 100Hz narrow square wave pulse + 400Hz wide square pulse wave. | 
		
			| Optical System | Monochomator | Single beam, Czerny-Turner design grating monochromator | 
		
			| Grating | 1800 l/mm | 
		
			| Focal length | 277mm | 
		
			| Blazed Wavelength | 250nm | 
		
			| Spectral Bandwidth | 0.1nm, 0.2nm, 0.4nm, 1.2nm, auto switch over | 
		
			| Flame Atomizer | Burner | 10cm single slot all-titanium burner    | 
		
			| Spray chamber | Corrosion resistant all-plastic spray chamber. | 
		
			| Nebulizer | High efficiency glass nebulizer with metal sleeve, sucking up rate: 6-7mL/min | 
		
			| Emission burner provided | 
		
			| Graphite Furnace | Temperature range | Room temperature~3000ºC | 
		
			| Heating rate |    2000℃/s | 
		
			| Graphite tube dimensions | 28mm (L) x 8mm (OD) | 
		
			| Characteristic mass | Cd≤0.8 ×10-12g, Cu≤5 ×10-12g, Mo≤1×10-11g | 
		
			| Precision | Cd≤3%, Cu≤3%, Mo≤4% | 
		
			| Detection and Data Processing System | Detector | R928 photomultiplier with high sensitivity and wide spectral range. | 
		
			| Software | Under Windows operating system | 
		
			| Analytical method | Working curve auto-fitting; standard addition method; automatic sensitivity correction; automatic calculation of concentration and content. | 
		
			| Repeat times | 1~99 times, automatic calculation of mean value, standard deviation and relative standard deviation. | 
		
			| Multi-task Functions | Sequential determination of multi-elements in the same sample. | 
		
			| Condition reading | With model function | 
		
			| Result printing | Measurement data and final analytical report printout, editing with Excel. | 
		
			| Standard RS-232 serial port communication | 
		
			| Graphite Furnace Autosampler | Sample tray capacity | 55 sample vessels and 5 reagent vessels | 
		
			| Vessel material | Polypropylene | 
		
			| Vessel volume | 3ml for sample vessel, 20ml for reagent vessel | 
		
			| Minimum sampling volume | 1μl | 
		
			| Repeatable sampling times | 1~99 times | 
		
			| Sampling system | Accurate dual pump system, with 100μl and 1ml injectors. | 
		
			| Characteristic Concentration and Detection Limit | Air-C2H2 flame Rich oxygen Air-C2H2 flame | Cu: Characteristic concentration ≤ 0.025 mg/L, Detection limit≤0.006mg/L; Ba: Characteristic concentration ≤ 0.22mg/L Al: Characteristic concentration ≤ 0.4mg/L | 
		
			| Function Expansion | Hydride vapor generator can be connected for hydride analysis. | 
		
			| Dimensions and Weight | Main unit | 107×49×58cm, 140kg | 
		
			| Graphite furnace | 42×42×46cm, 65kg | 
		
			| Autosampler | 40×29×29cm, 15kg |