| Main Specification | Wavelength range | 190-900nm | 
		
			| Wavelength accuracy | ±0.25nm | 
		
			| Resolution | Two spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley-peak energy ratio less than 30%. | 
		
			| Baseline stability | ≤0.004A/30min | 
		
			| Background correction | The D2 lamp background correction capability at 1A is better than 30 times. The S-H background correction capability at 1.8A is better than 30 times. (only for AE-WFX110A/120A) | 
		
			| Hollow Cathode Lamps | Lamp turret | 6-lamp turret (AE-WFX110A/120A), 4-lamp turret (AE-WFX130A) Auto-alignment, fully automated scan and peak-picking. | 
		
			| Lamp current adjustment | Automatic adjustment and display. Wide pulse current: 0~25mA,              Narrow pulse current: 0~10mA                 | 
		
			| Lamp power supply mode | 400Hz square wave pulse 100Hz Narrow square wave pulse + 400Hz wide square wave pulse (AE-WFX110A/120A) | 
		
			| Optical System | Monochomator | Single beam, Czerny-Turner design grating monochromator | 
		
			| Grating | 1800 l/mm | 
		
			| Focal length | 277mm | 
		
			| Blazed Wavelength | 250nm | 
		
			| Spectral Bandwidth | 0.1nm, 0.2nm, 0.4nm, 1.2nm automatic change. | 
		
			| Flame Atomizer | Burner | 10cm single slot all-titanium burner   | 
		
			| Spray chamber | Corrosion resistant all-plastic spray chamber. | 
		
			| Nebulizer | High efficiency glass nebulizer with metal sleeve, sucking up rate: 6-7mL/min | 
		
			| Emission burner | Provided with AE-WFX110A/120A | 
		
			| Graphite Furnace | Temperature range | Room temperature~3000ºC | 
		
			| Heating rate | 2000℃/s | 
		
			| Graphite tube dimensions | 28mm (L) x 8mm (OD) | 
		
			| Characteristic mass | Cd≤0.8 ×10-12g, Cu≤5 ×10-12g, Mo≤1×10-11g | 
		
			| Precision | Cd≤3%, Cu≤3%, Mo≤4% | 
		
			| Detection and Data  Processing System | Detector | R928 Photomultiplier with high sensitivity and wide spectral range. | 
		
			| Software | Windows operating system | 
		
			| Analytical method | Working curve auto-fitting; standard addition method; automatic sensitivity correction; automatic calculation of concentration and content. | 
		
			| Repeat times | Maximum 20 times of repeat measurement, automatic calculation of mean value, standard deviation and relative standard deviation. | 
		
			| Multi-task function | Sequential measurement of multi-elements in one sample | 
		
			| Condition reading | With model function | 
		
			| Result printing | Measurement data and final analytical report printout, editing with Excel. | 
		
			| Standard RS-232 serial port communication | 
		
			| Characteristic  Concentration and Detection Limit | Normal Air-C2H2 flame | Cu: Characteristic concentration≤0.025mg/L, Detection limit≤0.006mg/L; | 
		
			| Oxygen-rich Air-C2H2 flame | Ba:  Characteristic concentration ≤ 0.22mg/L Al:  Characteristic concentration ≤ 0.4mg/L (for WFX-110A) | 
		
			| Function Expansion | Hydride vapor generator can be connected for hydride analysis. | 
		
			| Dimensions and weight | 1020 (L) × 490(W) × 540 (H) mm (main unit), unpacked 80kg 420 (L) × 420 (W) × 460 (H) mm (graphite furnace), unpacked 50kg |