AA500 Atomic Absorption Spectrometer
Features:
Fully Automated
* The flame and graphite furnace is integrated into one instruments.The change over from one technique to another is carried out by simple key strokes within the software.
* The instruments have a motorised 8 hollow cathode lamp turret which allows the automatic positioning and optimization of each hollow cathode lamp by the software.
* The spectral bandwidth is automated and is available with choice of five slit sizes.
Advanced Graphite Furnace
* Horizontally heated platform graphite tube.The unique design of graphite furnace reduces the chemical
  interference effects and memory effects by uniformly heating the graphite electrode.
* The computer controlled heating program all.Iows the users to select the best heating program for analysis.
* The optical temperature during the atomization stage ensures the rapid heating and rapid analysis.This helps to  
  extend the life of graphite tube and enhance the analytical accuracy.
Ordering Information:
	
		
			| Order Code | Description | 
		
			| AE-AA500F | Single Beam, 190-900nm, With Flame. | 
		
			| AE-AA500G | Single Beam, 190-900nm, With Graphite Furnace. | 
		
			| AE-AA500FG | Single Beam, 190-900nm, With Flame & Graphite Furnace. | 
	
Technical Specifications:
	
		
			|  Optic System | Flame Analysis | 
		
			| Wavelength Range | 190nm-900nm | Sensitivity(Cu) | 2.0μg/ml Absorption>0.2 Abs | 
		
			|    Spectral Bandwidth |  0.1nm,02nm,0.4nm,1.0nm, 2.0nm (5steps,Automatic Changeover) |   Detection(Cu) | Cu<0.0004μg/ml(flame);Cd<0.4×10-12g(graphite furnace) Cu<0.7%(Air-Acetylene Analysis); Ba<1% (Nitrous Oxide Analysis) Cu<2%,Cd<2%(Graphite Furnace Analysis) | 
		
			| Monochromator | Czemy-Turner Configuration | Character Value | Cu<0.3×10-10g,Cd<0.3×10-12g | 
		
			| Wavelength Accuracy |    ±0.15nm | Burner Head | Titanium Alloy Burner | 
		
			| Wavelength Repeatability |    <0.05nm | Nebulizer | High-Efficiency Glass Nebulizer | 
		
			| Resolution  |   <30% or 0.2nm ±0.02nm | Atomization Chamber | Corrosion-Resistant Material | 
		
			| Baseline Stability | 0.005A/30min | Background Correction | Deuterium Lamp>×40,Self Reversal>×60 | 
		
			| Lamps | 8 lamps simultaneously lit | Character Concentration | <0.02μg/ml1%(Cu); <0.15mg/ml1%(Nitrous Oxide-Acetylene,Ba) | 
		
			| Power Supply | AC 220V,50Hz, 200W. | Position Adjustment | Automatic Changeover of Flame and Furnace Automatic setting of Optimum Height for Flame Burner | 
		
			| Packing Size & Weight | 1280x745x740mm, 160kg | Safety Functions |  Burner identification,Flame Sensor,Gas leak sensor, | 
	