Overall
Performance
|
Model
|
BIOBASE-DIAMOND
|
Type
|
discrete, fully automatic
|
Throughput
|
400 T/H
|
Analysis Method
|
End-point, Fixed-time,Rate
|
Programming Items
|
programmable for any test items
|
Certification
|
CE, FDA, ISO9001, ISO14001, ISO13485
|
Sample unit
|
Sample Settings
|
60/90 sample positions, including calibration, quality control and emergency.
|
Sample Volume
|
2ul ~50ul (0.1ul stepping)
|
Sample Probe
|
Having liquid level induction function.
|
Probe cleaning
|
Automatic washing both interior and exterior
|
Analysis Items
|
analyzing maximally 45 items at the same time
|
Reagent unit
|
Reagent Position
|
R1:45 R2:45 total 90 refrigerated reagent position
|
Reagent Volume
|
R1:25~300ul, R2:10~150ul (1ul stepping), supporting reagent from different brand
|
Temperature Control
|
continuous 2C~8Ccooling within 24 hours
|
Reagent Probe
|
Having liquid level induction function
|
Probe cleaning
|
Automatic washing both interior and exterior
|
Reaction unit
|
Reaction Time
|
10 minutes
|
Temperature
|
37C±0.1C
|
Cuvettes
|
120 reaction cuvettes (optical path length 6mm)
|
Mixer
|
after injecting reagent and sample, blending immediately
|
Cleaning
|
automatic water cleaning and drying system
|
System Adjustment
|
System record the blank value and get more accurate result by deducting blank base value of each cuvette
|
Optical System
|
Light Source
|
12V/30W halogen-tungsten lamp, 2000 hours
|
Receiver
|
8 precision photo receivers
|
Wavelength Range
|
340,405,450,510,546,578,630,700mm
|
Calibration and QC
|
Calibration
|
one-point,multi-point
|
Linear Range
|
0.000~2.500Abs
|
Quality Control
|
Inserting quality control at random.
|
Software
|
Emergency Function
|
emergency sample can be arranged at any time
|
Software
|
Windows XP or 7, sample and easy to operate
|
Data Processing
|
Having reaction curve detection and data display function in whole process.
|
Others
|
Power Supply
|
AC220V+22V/50Hz+1Hz
|
Dimension
|
1070mm×910mm×1230mm (W*D*H)
|
Weight
|
180kg
|