Infrared Oil Spectrophotometer
This instrument is used to test the oil content in water
Parameter:
Extremity stability: zero real time auto adjust
(eliminate influence from moving of baseline)
Detectable extremity: DL<=0.04mg/L
Repeatability: RSD<= 0.6%
(test 11 times for standard sample 30~40mg/L)
Line index: r>= 0.999
Waves range: 2941nm~4167nm
Absorption range: 0~2 (100%~1%T)
Waves’ accuracy: +/-2cm-1
Waves’ repeatability: +/-2cm-1
Basic test range: 0.1~80 mg/L
(4cm cuvette, oil content in extract liquor)
Lowest content: 0.001mg/L (oil content in water)
Highest content: 640000mg/L (oil content in water)
Scan speed: all spectrums scan, faster than 30 second
each time; no-disperse infrared way, 2 second per time.
Size: L400*W260*H150mm
N.W: 12 kgs
Power: 220+/-22V, 50+/-1 Hz, 50VA |