Atomic Absorption Spectrophotometer

| Technical Parameter of AAS6000 |
| MODEL |
PZ-AAS6000(acetylene fire) |
| Wavelength Range |
190-900nm |
| Optical System |
Total reflection achromatism |
| Monochromator Type |
C-T |
| Grating Number |
1800wires/mm |
| Spectral bandwidth |
0.1,0.2,0.4,0.7,1.4nm |
| Wavelength Accurancy |
±0.25nm |
| Wavelength Repeatability |
≤0.1nm |
| Element lamp |
8Lighthouse,1 lamp working,other 7 lamps preheat at most |
| Background correction System |
6000 Self-absorption & D2 lamp background correction |
| Communication interface |
USB2.0 |
Flame analysis |
Precision(Cu) |
0.01 |
| Feature Density(Cu) |
0.025ug/ml/1% |
| Detection limit(Cu) |
0.006ug/ml |
| Steady stability |
0.004Abs |
| Dynamic Stability |
0.005Abs |
| Acetylene flow control |
MFC Automatic flow control |

Following 33 elements can be tested:
Ag,As,Au,Bi,Ca,Cd,Co,Cr,Cs,Cu,Fe,Hg,In,Ir,K,Li,Mg,Mn,Na,Ni,Pb,Pd,Pt,Rb,Rh,Ru,Sb,Se,Sn,Sr,Te,Ti,Zn.