Purpose:
The U-50 series multi-parameter water-quality meter simultaneously measures and displays 11 parameters. Rugged, powerful and GPS-equipped, it is ideal for testing rivers, ground-water and waste-water.
Features:
• Read-outs in five languages (English, Japanese, Russian, Spanish, Portuguese)
• Simultaneous display of 11 parameters: pH, pH (mV), ORP, dissolved oxygen, conductivity, salinity, TDS, seawater specific gravity, temperature, turbidity and water depth
• U-53 turbidity probe meets US EPA requirements; resolution 0.01 NTU with automatic cleaning brush
• U-54 turbidity probe complies with EN ISO 7027; resolution 0.01 NTU
• Improved DO probe gives faster response and easier maintenance
Main Unit & Probe Specifications
|
Main Unit
|
|
|
Dimensions
|
115 mm W × 65 mm D × 280 mm H
|
|
Weight
|
approx. 800 g (1.76 lb)
|
|
LCD
|
320 × 240 back-lit monochrome
|
|
Data port / memory
|
USB, 10 000 data sets
|
|
Battery
|
4 × AA dry cells
|
|
Battery life
|
U-51/52/52G/54/54G ≈ 70 h (back-light off, 20 °C); U-53/53G ≈ 500 measurement cycles
|
|
Protection
|
JIS 7 (with cable attached)
|
|
Storage / operating temp.
|
–10 – 60 °C / –5 – 45 °C
|
|
Probe
|
|
|
Diameter × length
|
approx. 96 mm × 340 mm
|
|
Weight
|
approx. 1 800 g (3.97 lb)
|
|
Cable length
|
2 m std.; optional 10 m or 30 m
|
|
Measuring depth
|
0 – 30 m
|
|
Auto-cal.
|
pH 4 buffer (all models)
|
|
Wetted materials
|
PPS, glass, SIS316L, SUS304, FKM, PEEK, quartz, titanium, FEP membrane, POM
|
|
Protection
|
JIS 8
|
|
Turbidity wiper
|
built-in on U-53 & U-53G only
|
Measurement Performance
|
Parameter
|
Range
|
Accuracy
|
Repeatability
|
Resolution
|
|
Temperature
|
–10 – 55 °C
|
±0.3 °C
|
±0.1 °C
|
0.01 °C
|
|
pH
|
0 – 14
|
±0.1
|
±0.05
|
0.01
|
|
ORP
|
–2 000 – +2 000 mV
|
±15 mV
|
±5 mV
|
1 mV
|
|
Dissolved O₂
|
0 – 50 mg L⁻¹
|
±0.2 mg L⁻¹ (0-20); ±0.5 mg L⁻¹ (20-50)
|
±0.1 mg L⁻¹
|
0.01 mg L⁻¹
|
|
Salinity / Conductivity
|
0 – 70 ppt
|
±3 ppt
|
±1 ppt
|
0.1 ppt
|
|
Depth
|
0 – 30 m
|
±0.3 m
|
±1 % FS
|
0.05 m
|
|
Turbidity (LED) U-51/52/52G
|
0 – 800 NTU
|
±3-5 % FS or ±1 NTU
|
±3-5 % FS or ±0.5 NTU
|
0.1 NTU (0-100); 1 NTU (100-800)
|
|
Turbidity (LED) U-53/53G/54/54G
|
0 – 1 000 NTU
|
same as above
|
same as above
|
same as above
|
|
TDS
|
0 – 100 g L⁻¹
|
±5 g L⁻¹
|
±2 g L⁻¹
|
0.1 g L⁻¹
|
|
Seawater specific gravity
|
0 – 50 δt
|
±5 δt
|
±2 δt
|
0.1 δt
|
Notes
*Battery life measured with alkaline AA cells, display temp. ≥ 20 °C, back-light off.
*Accuracy for turbidity & conductivity after 4-point calibration; other items after 2-point calibration with standard solutions.
*Repeatability values refer to measurements in standard solutions.
Model Comparison (✓ = included)
|
Parameter
|
U-51
|
U-52
|
U-52G
|
U-53
|
U-53G
|
U-54
|
U-54G
|
|
pH
|
✓
|
✓
|
✓
|
✓
|
✓
|
✓
|
✓
|
|
ORP
|
✓
|
✓
|
✓
|
✓
|
✓
|
✓
|
✓
|
|
Dissolved oxygen
|
✓
|
✓
|
✓
|
✓
|
✓
|
✓
|
✓
|
|
Conductivity
|
✓
|
✓
|
✓
|
✓
|
✓
|
✓
|
✓
|
|
Salinity
|
✓
|
✓
|
✓
|
✓
|
✓
|
✓
|
✓
|
|
TDS
|
✓
|
✓
|
✓
|
✓
|
✓
|
✓
|
✓
|
|
Seawater specific gravity
|
✓
|
✓
|
✓
|
✓
|
✓
|
✓
|
✓
|
|
Temperature
|
✓
|
✓
|
✓
|
✓
|
✓
|
✓
|
✓
|
|
Turbidity (LED)
|
—
|
✓
|
✓
|
—
|
—
|
✓
|
✓
|
|
Turbidity (tungsten)
|
—
|
—
|
—
|
✓
|
✓
|
—
|
—
|
|
Water depth
|
—
|
—
|
✓
|
✓
|
✓
|
✓
|
✓
|
|
GPS
|
—
|
—
|
✓
|
—
|
✓
|
—
|
✓
|
*U-54/G (2 m cable) has no depth sensor.