It is an analytical instrument based on emission spectrum principle, using the thermal energy of flame to make the certain element motivate luminescence to analyze the content of certain element in the material ,widely used in the analysis and determination of agricultural fertilizer, soil, cement, ceramics and silicate industry and pathological studies of health care etc.
Features:
* USB Interface
* LPG as burning gas
* Direct concentration read-out
* Measuring range adjustable
* 7 inch LCD touch screen
* Automatic Calculation of correlation coefficient
* Pre-selection of flame sizes
* Flameout protection device
* Simultaneous detection and display of all five elements
Model
|
|
FP6540
|
FP6440
|
FP6432
|
FP6431
|
FP6430
|
Screen
|
7” Color touch LCD
|
Display
|
Concentration/Flame Intense
|
Reading Value
|
0.000~999.9/0~65535
|
Measured Elements
|
K; Na; Li; Ca; Ba
|
K; Na; Li; Ca
|
K; Na; Ba
|
K; Na; Ca
|
K; Na; Li
|
Channels
|
5
|
4
|
3
|
3
|
3
|
Detection Limit
Mmol/L
|
K
|
≤0.004
|
Na
|
≤0.008
|
Li
|
≤0.015
|
——
|
——
|
≤0.015
|
Ca
|
≤0.050
|
——
|
≤0.050
|
——
|
Ba
|
≤0.044
|
——
|
≤0.044
|
——
|
——
|
Linearity
mmol/L
|
K
|
≤0.005
|
Na
|
≤0.03
|
Li
|
≤0.021
|
——
|
——
|
≤0.021
|
Ca
|
≤0.075
|
——
|
≤0.075
|
——
|
Ba
|
≤0.066
|
——
|
≤0.066
|
——
|
——
|
Response Time
|
|
<8s
|
小吸喷量
|
<6ml/min
|
Stability
|
Continuous injection of standard solution, relative significant change of reading value less or equal with 3% in 15s;
Measure once per minute for a total of 6 times, significant change of reading value less or equal with ≤15%.
|
Repeatability
|
≤3%
|
Curve
|
Support
|
Printer
|
Optional
|
Printer type
|
Thermal Printer/Wireless Bluetooth printer
|
USB Interface
|
Yes
|
Connect PC
|
Support
|
bio-equip.cn