| Model |
LYX-50 |
LYX-150 |
| Type |
Six-sided heating + breeze circulation + air jacket |
| Temperature control range |
RT+5℃-55℃ |
| Temperature resolution |
±0.1℃ |
| Constant temperature fluctuation |
±0.1℃ |
| Temperature uniformity |
0.35℃ |
| CO2 control range |
0-20% |
| CO2 distribution accuracy |
±3% |
| Humidity in the box |
≥90% |
| Interior |
SUS304Cu mirror stainless steel |
| Exterior |
cold-rolled steel plate, surface chemical-resistant coating |
| Thermal insulation material |
polyurethane |
| Heater |
Silicone heating patch |
| CO2 inlet |
φ8mm |
| CO2 filter |
DUF imported |
| Partition |
SUS304Cu mirror stainless steel |
| Temperature control method |
PID |
| CO2 concentration control method |
PID |
| Temperature setting method |
touch 6 button to set |
| Temperature expression method |
3 digits |
| CO2 concentration indication method |
3 digits |
| Timer |
0-9999 minutes (with timing waiting function) |
| Operation function |
fixed value operation, timing operation, automatic stop, quick stop start |
| Additional functions |
deviation correction, menu button lock, power failure compensation, power failure memory |
| Temperature sensor |
Pt100 |
| CO2 concentration sensor |
imported |
| Water level sensor |
imported |
| Additional function |
Deviation correction, menu key lock, power failure compensation, power failure memory |
| Safety device |
Over-rising alarm, menu lock, CO2 abnormal alarm, low water level alarm, power failure protection |
| Chamber size (mm) |
350x380x380 |
500x500x600 |
| Dimension (mm) |
465x480x580 |
615x700x800 |
| Package size (mm) |
640x734x711 |
790x850x931 |
| Volume |
50L |
150L |
| Tray loading weight |
15kg |
15kg |
| Tray No. |
3 |
3 |
| Max. tray No. |
6 |
6 |
| Tray distance |
35mm |
35mm |
| Weight (Net/Gross) |
46/55kg |
76/88kg |
| Rated power |
0.18 kw |
0.33kw |
| Power |
AC220V / 0.8A |
| Optional |
Separator, multi-gas culture, automatic switching of dual gas cylinders |