Screen Type: High-resolution display
Screen Size: 7 inches (17.7 cm) diagonally
Sweep Speed: 25mm/sec
Information: Heart Rate Lead/Pads Alarm On/Off SpO2 AED
Functions and Prompts Alarm Selection and Limits Delivered Energy.
5 Languages available: English.Russian.Spanesh.Chinese.Turkish
Defibrillator
Waveform: Biphasic
Charge Time: Less than 7 seconds with a new fully charged battery.
Energy Display: Monitor display indicates both selected and delivered energy.
Charge Prompt Type: Voice and visual prompts.
Electrode Impedance Measurement Range: 0-250 ohms.
ECG Monitoring
Patient Connection: 3-lead ECG cable or 5-lead ECG cable paddles.
Lead Selection: Displayed on monitor paddles I II III AVR AVL AVF V.
ECG Size: 0.25 0.5 1 1.5 2 4 cm/mV display on monitor.
Heart Rate: 20-300BPM.
Heart Rate Alarm: On/Off displayed on monitor user-selectable.
Smart Alarms: Beeper/voice prompts indicate shock able rhythm.
Battery
Type: Rechargeable Ni-MH battery 12V.
Operating Time: For a new fully charged battery: 60 defibrillator discharges at maximum energy or 3 hours minimum of continuous ECG monitoring. Additional parameters will effect operating time with different functions
SpO2 Module (Options):
Measurement Range: 30~100% ±2% between 80%~90% others ±5%
Alarm Range: User set high limit and low limit.
Alarm Accuracy: ±10% within setting values.
Alarm time Accuracy: less than 12 sec.
1-Meditech Pulse Oximeter
2-Nellcor Pulse Oximeter
Recorder
Paper: 50mm thermal.
Speed: 12.5mm/sec 25mm/sec 50mm/sec. User-selectable 6-second delay.
Printing Method: High-resolution thermal print head.
Print-out Modes: Manual or automatic user-configurable.
On/Off Control: Front panel and paddle.
Automatic Function: 9-seconds recording initiated by alarm activation or defibrillator charge or defibrillator discharge.
AED Mode
AED Function: uto analyze and charge x3 with programmable auto energy level selection screen prompts and voice prompts.
Shockable Rhythms: Ventricular fibrillation with amplitude >=200uV ventricular tachycardia with rates >=140bpm and QRS complex wave duration >=140ms.
Charge Control: Control on device front panel press key on paddle.
Prompts: Voice and visual prompts.
Manual Mode
Energy Selection: Selectable at 2 5 7 10 20 30 50 70 100 150 200 300 360 joules.
Synchronized Mode: Synchronizes defibrillator pulse to patient’s
R-wave. ”SYNC” message displayed on monitor.

bio-equip.cn