Wavelength range: 190 to 1100 nm.
Wavelength accuracy: ± 0.3nm
Wavelength repeatability: 0.2 nm
Photometric Scope: -3.000 A ~ 3.000A
Photometric accuracy: ± 0.3% T (0 to 100% T)
Photometric repeatability: 0.001 Abs (0 ~ 0.5Abs)
- Stability: 0.0004 Abs / h
The baseline flatness: ± 0.001Abs
Stray light: ≤ 0.05% T (220nm, 340nm)
Noise: <± 0.0003Abs
Scanning speed: slow in the fast
Spectral bandwidth: 2 nm
Voltage: 220V 50Hz
Uv1900 series spectrophotometer in accordance with the standards of the Chinese production of the chemical spectrophotometer, in the application of environmental control (air-soil water), water safety, as well as the manufacturers of raw materials and finished products (food chemical pharmaceutical steel petrochemical environmental), and the analysis of the needs of other areas.
* Newly designed excellent optical system, high-performance holographic blazed grating to ensure that the equipment and low stray light.
* Dual-beam metering systems to meet the circuit design of the measurement and control system, so that machines with a high degree of stability
* The very low noise
* Automatic control systems, advanced design concepts to ensure that machines with high reliability and high stability
* Removable structure of the sample chamber design, easy replacement of different attachments, to meet the different needs analysis
* Open-source spacious rooms designed to enable more convenient lamp replacement source
* Selection of key components are imported devices, ensuring the performance of high-reliability equipment
* Windows environment in the development of the English operating software, and has a number of patented technologies, provide a wealth of unique functional analysis