Login    Join free Add to favorites    中文
Home >Products> Pumps >Peristaltic pump> Syringe pump
Syringe pump
Syringe pump
Origin of place China
Model LSP01-1A
Supplier LONGER Precision Pump Co.Ltd
Price
Hits 1962
Updated 11/12/2019
  • Product Detail
  • Company Profile

 

Instruction
This pump is a single channel syringe pump which only has infusion mode. The acceptable syringe specification is from 10 μL to 60 mL. Suitable for high accuracy and small flow rate liquid transferring.
Specifications
Syringe pump LSP01-1A
Advance per Mircostep 1/16step: 0.156 μm
Infusion Volumn per Microstep (60 mL BD Syringe) 0.0867 μL 
Max. Linear Rate 65 mm/min
Min. Linear Rate 5 μm/min
Flow Rate  0.831 nL/min -  54.155 ml/min
Working Mode     Infusion
Max. No. of Syringes 1
Syringe Size 10 μL - 60 mL
Linear Force > 9 kg
Max. Step Rate 6933  (1/16step)/ sec
Min. Step Rate 16  (1/16 step) /30 sec
Accuracy ≤ 0.5% error in the condition of   > 30% of max. infusion distance
Setting Mode Rotary coded switch and membrane keypad
Display 128 x 64 graphic LCD
Power AC 100 - 240 V
Operating Condition Temperature  5 °C - 40 °C       Relative humidity < 80 %
Dimensions 280 × 210 × 140 mm
Weights 3.6 kg
Functions and Features
- Syringe Selection: The syringe can be selected in the manufacturer table which includes manufacturer, material and size or input the inner diameter of the syringe barrel directly
- Easy to operate: Combining big screen LCD display with rotary coded switch and membrane keypad makes the operation simple and prompt.
- Working Mode: Infusion
- Memory Function:
1. The parameters are saved in EEPROM. The parameters don't need to be reset when power returns after an interruption
2. In flow rate mode, the pump remains running or stop according to the setting parameters when power returns after an interruption
- Protection Function: The pump will stall and give an alarm when the drive structure of the pump is blocked
- Communication Function: Realize computer control through RS485 communication interface
- External Control Function: Input/output control
- Calibration Function: Acquire accurate volume through calibration
- Syringe Protection: Adjust syringe rest to prevent syringe from damaging
Parameters
Hamilton Syringe (mL)  Diameter(mm) LSP01-1A
Ref. Flow Rate(*/min) Linear Rate(*/min)
50 uL 1.03 0.004 L - 0.054 mL    5 μm - 65mm    
10 14.57 0.834 μL - 10.837 mL
60 32.57 4.166 μL - 54.155 mL
bio-equip.cn
Request Infomation

* Name:
Title:
* Tel:
Fax:
* E-mail: