Prolactin (PRL) Elisa Test Kit for Diagnostic Use
Intended use
Immunoassay for the in vitro quantitative determination of prolactin in human serum.
Summary
Prolactin is synthesized in the anterior pituitary and is secreted in episodes. The hormone is made up of 198 amino acids and has a molecular weight of approx. 22‐23 kD. Prolactin appears in serum in three different forms. The biologically and immunologically active monomeric (“little”) form predominates (approx. 80 %), 5‐20 % is present as the biologically inactive dimeric (“big”) form and 0.5‐5 % is present as the tetrameric (“big‐big”) form having low biological activity.(1) The target organ for prolactin is the mammary gland, the development and differentiation of which is promoted by this hormone. High concentrations of prolactin have an inhibiting action on steroidogenesis of the ovaries and on hypophyseal gonadotropin production and secretion.During pregnancy the concentration of prolactin rises under the influence of elevated estrogen and progesterone production. The stimulating action of prolactin on the mammary gland leads post partum to lactation.
Hyperprolactinemia (in men and women) is the main cause of fertility disorders. The determination of prolactin is utilized in the diagnosis of anovular cycles, hyperprolactinemic amenorrhea and galactorrhea, gynecomastia and azoo-spermia. Prolactin is also determined when breast cancer and pituitary tumors are suspected.(2)
Test principle
Sandwich principle. Total duration of assay: 80 minutes.
- Sample, Anti-Prolactin coated microwells and enzyme labeled Anti-Prolactin are combined.
- During the incubation, prolactin presents in the sample is allowed to react simultaneously with the two antibodies, resulting in the prolactin molecules being sandwiched between the solid phase and enzyme-linked antibodies.
- After washing, a complex is generated between the solid phase, the prolactin within the sample and enzyme-linked antibodies by immunological reactions.
- Substrate solutionis then added and catalyzed by this complex, resulting in a chromogenic reaction. The resulting chromogenic reaction is measured as absorbance.
- The absorbance is proportional to the amount of PRL in the sample.
Reagents
Materials provided
- Coated Microplate, 8 x 12 strips, 96 wells, pre-coated with mouse monoclonal Anti-Prolactin.
- Calibrators, 6 vials, 1 ml each, ready to use; Concentrations: 0(A), 5(B), 10(C), 20(D), 50(E) and 100(F) ng/mL.
- Enzyme Conjugate, 1 vial, 11 mL of HRP(horseradish peroxidase) labeled mouse monoclonal Anti-prolactin in Tris-NaCl buffer containing BSA (bovine serum albumin). Contains 0.1% ProClin300® preservative.
- Substrate, 1 vial, 11ml, ready to use, (tetramethylbenzidine) TMB.
- Stop Solution, 1 vial, 6.0 ml of 1 mol/L sulfuric acid.
- Wash Solution Concentrate, 1 vial, 25 ml (40X concentrated), PBS-Tween wash solution.
- IFU, 1 copy.
- Plate Lid: 1 piece.
Marerials required (but not provided)
Microplate reader with 450nm and 620nm wavelength absorbent capability.
- Microplate washer.
- Incubator.
- Plate shaker.
- Micropipettes and multichannel micropipettes delivering 50μl with a precision of better than 1.5%.
- Absorbent paper.
- Distilled water
Precautions and warnings
- For in vitro diagnostic use only. For professional use only.
- All products that contain human serum or plasma have been found to be non-reactive for HBsAg, HCV and HIVI/II. But all products should be reared as potential biohazards in use and for disposal.
- Mix the sample in the wells thoroughly by shaking and eliminate the bubbles.
- Conduct the assay away from bad ambient conditions. e.g. ambient air containing high concentration corrosive gas such as sodium hypochlorite acid, alkaline, acetaldehyde and so on, or containing dust.
- Wash the wells completely. Each well must be fully injected with wash solution. The strength of injection, however, is not supposed to be too intense to avoid overflow. In each wash cycle, dry the liquids in each well. Strike the microplate onto absorbent paper to remove residual water droplets. It is recommended to wash the microplate with an automated microplate strip washer.
- Failure to remove adhering solution adequately in the aspiration or decantation wash step(s) may result in poor replication and spurious results.
- Do not use reagents beyond the labeled expiry date.
- Do not mix or use components from kits with different batch codes.
- If more than one plate is used, it is recommended to repeat the calibration curve.
- It is important that the time of reaction in each well is held constant to achieve reproducible results.
- Ensure that the bottom of the plate is clean and dry.
- Ensure that no bubbles are present on the surface of the liquid before reading the plate.
- Thesubstrate and stop solution should be added in the same sequence to eliminate any time deviation during reaction.
Storage
- Store at 2-8℃.
- Place unused wells in the zip-lock aluminum foiled pouch and return to 2-8 °C, under which conditions the wells will remain stable for 2 months, or until the labeled expiry date, whichever is earlier.
- Seal and return unused calibrators to 2-8 °C, under which conditions the stability will be retained for 1 month, for longer use, store opened calibrators in aliquots and freeze at -20 °C. Avoid multiple freeze-thaw cycles.
- Seal and return all the other unused reagents to 2-8 °C, under which conditions the stability will be retained for 2 months, or until the labeled expiry date, whichever is earlier.
Specimen collection and preparation
- Collect serum samples in accordance with correct medical practices.
- Cap and store the samples at 18-25 °C for no more than 8 hours. Stable for 7 days at 2-8 °C, and 1 month at -20 °C. Recovery within 90-110 % of serum value or slope 0.9-1.1. Freeze only once.
- The sample types listed were tested with a selection of sample collection tubes that were commercially available at the time of testing, i.e. not all available tubes of all manufacturers were tested. Sample collection systems from various manufacturers may contain differing materials which could affectthe test results in some cases. When processing samples in primary tubes (sample collection systems), follow the instructions of the tube manufacturer. Centrifuge samples containing precipitates before performing theassay. Do not use heat-inactivated samples. Do not use samplesand controls stabilized with azide.
- Ensure the patients’ samples, calibrators, and controls are at ambient temperature (18-25 °C) before measurement.
- Sediments and suspended solids in samples may interfere with the test result which should be removed by centrifugation. Ensure that complete clot formation in serum samples has taken place prior to centrifugation. Some samples, especially those from patients receiving anticoagulant or thrombolytic therapy, may exhibit increased clotting time. If the sample is centrifuged before a complete clot forms, the presence of fibrin may cause erroneous results. Be sure that the samples are not decayed prior to use.
- Avoid grossly hemolytic, lipemic or turbid samples.
- Note that interfering levels of fibrin may be present in samples that do not have obvious or visible particulate matter.
- If proper sample collection and preparation cannot be verified, or if samples have been disrupted due to transportation or sample handling, an additional centrifugation step is recommended. Centrifugation conditions should be sufficient to remove particulate matter.
- Ensre the patients’ samples, calibrators, and controls are at ambient temperature (18-25 °C) before measurement. Mix all reagents through gently inverting prior to use.
- Adjust the incubator to 37 °C.
- Prepare wash solution concentrate before measurement. Stable for 2 months at ambient temperature.
- Don’s use Substrate if it looks blue.
- Don’t use reagents that are contaminated or have bacteria growth.
Contact Rebecca Yan
Skype: span.biotech
Email: rebecca@spanbio.com
Tel: +86(755)89589611
Fax: +86(755)89580096
Web:
www.spanbiotech.com
bio-equip.cn
Span Biotech Ltd. is a research based company for rapid tests, with strong support from National Key Laboratory of Technology Projects of 10th and 11th Five-Year Plan and Faculty of Life Sciences of HuBei University. SpanBio also housed a R&D team that is developing gene recombination, cell cultivation and protein purification techniques. SpanBio pays strict attention on rapid tests for human being, animal diseases and food safety detection. It provides a number of customized services to professional distributors and partnering affiliates with excellent quality, competitive prices and super service.
Efficiently structured to provide lasting and economic health care solutions with export ready goods, our main products cover the following immunodiagnostic rapid test kits:
Rapid Tests for Human Use :
1) Fertility: Pregnancy HCG, ovulation LH, menopause FSH
2) Infectious diseases: Hepatitis B, HCV, HIV
3) Drug abuse: AMP, BAR, BZO, COC, mAMP, MOP, MTD, OPI, PCP, TCA, THC
4) Tumor markers: CEA, PSA, AFP, FOB
5) STDs(Sexually Transmitted Diseases): Gonorrhea, syphilis, chlamydia
6) New products: Malaria, H. Pylori, Multi-Hepatis B Panel (HBsAg, Anti-HBs, HBeAg, Anti-HBe, Anti-HBc), Blood Glucose Test, Multi-drug of Abuse Panel, Avian (Bird) Flu Rapid Test, etc.
Rapid Tests for Veterinary :
1) Bovine: Cow Pregnancy , Cow Ovulation, Bovine Brucella, Bovine Tuberculosis ,Food and Mouth Disease NSP Ab Test
2) Canine: Canine Distemper, Canine Parvovirus, Canine Coronavirus
3) Feline : Feline Leukemia Virus Ag Rapid Test ,Feline Panleucopenia Virus Ag Test,Feline Immunodeficiency Virus Ab Test
Rapid Tests for Food Safety :
1)For Antibiotics Residue : Nitrofurans (AOZ,AMOZ, AHD, SEM), Streptomycin , Sulfonamides , Chloramphenicol , Quinolone , Beta-Lactam&Tetracycline Combo (2 in 1), β-Lactams & Streptomycin & Chloramphenicol & Tetracyclines Combo (4 in 1), Beta-Lactam, Tetracycline test
2) For Beta-Agonist Residue : Ractopamine, Clenbuterol ,Salbutamol
|