Login    Join free Add to favorites    中文
Home >Products> Other Lab Equipments >Food & Beverage> 5E Series C/H/N Elemental Analyzer
5E Series C/H/N Elemental Analyzer
Origin of place China
Model 5E-CHN2200
Supplier Changsha Kaiyuan Instruments Co., Ltd
Price $67200
Hits 1848
Updated 8/15/2024
  • Product Detail
  • Company Profile

5E Series C/H/N Elemental Analyzer is used to determine carbon, hydrogen, nitrogen content in solid and liquid material, such as coal, coke, oil, petroleum, biomass, fertilizer, plastic, food, hydrocarbons and plant tissue, leaves and tobacco, which is widely applied in power plants, coal mines, metallurgy, chemical industry, commercial inspection, scientific research, food industry, education etc.

Features:

Maximum Efficiency
1. High throughput: standard auto loader for 35 samples per layer, stackable to 4 layers available.
2. Dual-stage furnace system with pure oxygen flow to ensure the complete combustion of all samples.

Good Environment Adaptability 
1. Optimum gas circuit provides good gas tightness of the system.
2. O-ring free from heat resource.

Minimum Consumption
1. Independent detectors to determine C, H, N respectively. Analysis of CH mode and CHN mode can be chosen on software. (For 5E-CHN2200)
2. Saving time, gas and reagent: only 5.5ml blended gas needed to be analyzed.

Unattended Operation
The operator is limited to just adding sample to auto sample loader. Then the instrument will finish the test, cool down and shut off automatically.

Specifications:

Model 5E-CHN2200 Elemental Analyzer
Conforms to Method AS 1038.6.4, ASTM D5373, ASTM D5291,
ISO 16634, ISO 16948, ISO 29541, EN 15407,EN ISO 16948:2015-0, 
GB/T 30728, GB/T 30733, UNI 15104, UNI CEN/TS 15407
Analysis Time 4-6 mins, depending on sample combustion conditions
Sampler Loader Stackable auto loader, up to 140 samples by 4 layers
Repeatability Carbon(Cad)≤0.45%, Hydrogen(Had)≤0.10%, Nitrogen(Nad)≤0.05%
Sample Mass Up to 1000mg, depending on sample matrix
Temp. Resolution 1℃
Gas Required* Helium, 99.99%, 0.25± 0.01 Mpa
Oxygen, 99.5%, 0.25± 0.01 Mpa
Nitrogen or compressed air, 0.25± 0.01 Mpa
Consumption Helium 200ml/min
Measurement Range Carbon: 0.02mg-150mg
Hydrogen: 0.1mg-12mg
Nitrogen: 0.04mg-50mg
Furnace Type Resistance furnace, max. temp 1050℃
Power Supply Single phase, AC220V ±10% , 50/60Hz, 5.5kW
Net Weight 110kg
Dimensions(L×W×H) 690mm×750mm×720mm
*Test Condition: If N≤0.5%, 99.995% high purity oxygen is required.
bio-equip.cn
Request Infomation

* Name:
Title:
* Tel:
Fax:
* E-mail:
Postcode:
Institution/Company:
Address:
* Country:
Request infomation:
yes no
Request Quotation:
yes no
* Message:
Validated Code:
refresh
I agree to share my inquiry to the other matching suppliers.



Copyright(C) 2006-2025 Bio-Equip    E-mail:web@bio-equip.cn   沪ICP备06040519号