Description 

GP-C18 uses full coverage bonded silica packing, which provides exceptionally high stability. 

The unique mono-functional bonding chemistry of GP-C18 avoids the formation of multiple C18 layers. Such uniform stationary phase 

allows the separation 

to achieve high selectivity and high efficiency.

Packing 


Silica:

Spherical, high purity (<10 ppm metals)

Pore size:

120Å

Particle size:

1.8, 2.2, 3, 4, 5, 7, and 10 µm

Pore volume:

1.0 ml/g

Surface area:

300 m²/g

Phase structure:

Monomeric and fully endcapped

% Carbon:

17%

Characteristic

 

· Highly controlled chemistry of monolayer formation and end-capping

· Extremely high column-to-column reproducibility

· High selectivity and efficiency for separations

· Suitable for separations of acidic, neutral and basic organic compounds, 

as well as pharmaceuticals, peptides, and others

· Recommended for separations in organic or mixed organic/aqueous mobile phases.

 

 

 

Stability

SepaxGP-C18 uses full coverage bonded silica packing, which provides exceptionally 

high stability. Figure below

 shows the highly reproducible retention time for three standard compounds: aniline,

 anisole and toluene

 after 13,000 column volume runs in a mobile phase of 85% methanol and 15% water. 

Such high stability allows SepaxGP-C18 extremely suitable for validation of various analytes.

Selectivity for GP-C18 and the Competitors' C18 Phases

Selectivity is very important for detection of the components of low abundance and

 impurities from 

the analysis of pharmaceuticals in the real applications. The comparative tests shown

 below from the method development of 

a drug molecule clearly demonstrated that GP-C18 is among the best products for high selectivity and sensitivity in detecting low 

abundant components and the impurities.

(Courtesy of Dr. Fang Liu, Novartis)

Conditions

Column:

Inertsil ODS-3, 3 µm (Varian)
Xterra RP18, 3.5 µm (Waters)
Suplex PKB-100, 5 µm (Supelco)
Sepax GP-C18, 3 µm

Mobile phase:

A: 99% Boric buffer (5.0 mM, pH 7.0)/1% Acetonitrile
B: 40% Boric buffer (5.0 mM, pH 7.0)/60% Acetonitrile

Gradient:

0-3 min (100% A); 3-38min (0-100%B); 38-43 min (100% B)

Column Temperature:

50 °C

Flow rate:

1.0 mL/min

Detection wavelength:

210 nm

Sample:

A real pharmaceutical sample (undisclosed)

bio-equip.cn
Request Infomation

* Name:
Title:
* Tel:
Fax:
* E-mail:
Postcode:
Institution/Company:
Address:
* Country:
Request infomation:
yes no
Request Quotation:
yes no
* Message:
Validated Code:
refresh
I agree to share my inquiry to the other matching suppliers.



Recently viewed products
Copyright(C) 2006-2025 Bio-Equip    E-mail:web@bio-equip.cn   沪ICP备06040519号