1.Product Introduction

HF6000 series blood coagulation analyzer (IVD) is suitable for all levels of hospitals.  It is easy to operate and highly accurate, with low consumption of reagents, reasonable price, and good data management capability.There are two types of HF6000 series coagulation analyzer:HF6000-2 coagulation analyzer (dual-channel), HF6000-4 coagulation analyzer (four-channel).

2.Measurement Principle

The measurement principle of the analyzer is photoelectric magnetic beads method, that is:using photoelectric to test movements of magnetic beads, in order to determine the freezing point of the blood,and measure the time of blood freezing.This method is free from the interference of Jaundice, Hemolytic,Chyle and other samples.

3.Advantages:

a.Blue screen with LED backlight, 5.7 inch, 320×240 graphic LCD display module

b.Thermal Micro-Printer

c.Power-off protection (data can be kept 10 years)

d.Can store data of 999 patients, and more than ten thousands test records

e.Real time display (year, month, day, week, hour, minute, second

f.Can be connected to computer via RS232 serial port for data exchange (long-distance data downloading, item change and interface switch)

g.Auto-save test result, and you can print real-time report and complete report

k.Operation instructions and simple dialogues with the instrument are available

4.Technical Parameters

      Semi automatic coagulation machine blood coagulation analyzer

Analysis methods

photo-electric detection magnetic beads method, free from the interference of jaundice, hemolysis, chyle, turbidity and bubble

Two instruments

dual channel HF6000-2, four channel HF6000-4

Preheat timer

preheat timer system,prompt automatically

The scope of detection

the scope of the original fiber protein detection, the detection results can be high or low plasma abnormal raw fiber protein.

Preheat temperature

37°C±1°C

Preheat time

15-30 minutes

Results saving

automatically, real time reports and comprehensive reports printed

Main test items

Prothrombin Time (PT),Activated Partial Thromboplastin Time( APTT),Fibrinogen (FIB),Thrombin time(TT)

Power off protection

10000 records that can store as long as 10 years

Open reagent

small sample consumption

Display

5.7 inches reversible LCD screen

Voltage

AC 220V±22V,50Hz±1Hz

Input voltage

≥170VA

working temperature

10°C-30°C

relative humidity

30%-80%

Dimension

490nm×435nm×280nm

Weight

7.5kg

bio-equip.cn
Request Infomation

* Name:
Title:
* Tel:
Fax:
* E-mail:
Postcode:
Institution/Company:
Address:
* Country:
Request infomation:
yes no
Request Quotation:
yes no
* Message:
Validated Code:
refresh
I agree to share my inquiry to the other matching suppliers.



Copyright(C) 2006-2025 Bio-Equip    E-mail:web@bio-equip.cn   沪ICP备06040519号