 Display: Large LCD 240×128 

Detection method : Absorbance

Microplate types: 96-well plate(U,V or flat-bottom)

Measuring system: 8-Channel optical ststem

Reading speed: 12s single wavelength, 30s double wavelength

Light source: Tungsten Halogen, 12V/20W, >2000hours 

405nm,450nm: 492nm,630nm,4free positions for additional filters 

Onboard software: 42user-programmable protocols
Reading range: 0.001-3.500Abs
Measurement range: 0.001-2.500Abs
Photometric accuracy: <1% or ± 0.001Abs 

Wavelength accuracy: <2nm 

Repeatability: <1% 

Stability: <0.005Abs/10min 

Linearity: <2.0% 

Shaking:linear shaking; Low,medium,high 

Input:Touch screen 

Power supply: AC220V/110±10%,50/60Hz 

Operating environment: Temperature:10°C ~35°C 
                                      Humidity:15~85% 

8-Channel optical measuring system
Optional single/double wavelength analysis
Variable speeds linear shaking
Cutoff and call criteria