Login    Join free Add to favorites    中文
Home >Products> Medical Lab Instruments >Microplate Readers / Plates Washers> Microplate Reader
Microplate Reader
Microplate Reader
Origin of place United States
Model RNE90002
Supplier REAGEN kits
Price By e-mail
Hits 1309
Updated 5/3/2020
  • Product Detail
  • Company Profile

Weight:

8.1kg

Overall dimensions:

430mm(L)×335mm(W)×175mm(H)

Power:

a.c.110V~220V,50-60Hz

Fuses:

T2AL250V

Work Environment:

temperature 10℃-30℃;Humidity ≤70%

Store Environment:

-20℃-55℃;Humidity ≤93%

Standard Wavelength:

405,450,492,630nm,Ten optical fibers can be configured within 400 to 800 at a time;

Instrument sensitivity:            

no less than 0.01L/mg.

Instrument channel difference:

less than 0.01A.

Range of operating wavelength:

400-800nm;

 

Detection range:

0.000-4.000A;

Resolution ratio:

0.001A;

Plate reading velocity:

5s of single waveform, 10s of dual wavelength;

Warm up time:

10 minute

Interface:

RS-232C serial interface, USB interface

Display:

5.7″LCD display(320×240 discernibility,256 gray scale)

Input:

Touch panel and pen, 

Technical Specifications

1) Wavelength characteristics: requirements on accurate wavelength and half wave width of optical fibers for instrument see the table below.

Unit: nm

Accurate wavelength

Half wave width

±3.0

≤12

2) Accurate absorbance and linear error: measurement accuracy and linear error of instrument shall comply with the provisions as described in the table below:

Absorbance range

Accuracy

Linear error

0.000-2.000

±0.01

±1.0%

>2.000-3.000

±0.05

±2.0%

>3.000-4.000

±0.15

±5.0%

3) Repeatability of measurement of instrument: no greater than 0.5%.

4) Instrument stability:

a) Absorbance varies shall be no greater than ±0.005A within 10min due to drifting.

b) If the power supply voltage of 220V varies by ±10%, the resulting variation of absorbance is no greater than ±0.005A.

bio-equip.cn
Request Infomation

* Name:
Title:
* Tel:
Fax:
* E-mail:
Postcode:
Institution/Company:
Address:
* Country:
Request infomation:
yes no
Request Quotation:
yes no
* Message:
Validated Code:
refresh
I agree to share my inquiry to the other matching suppliers.