Description
Lead Fluid MLP series products are mainly used for high-precision filling of various aqueous solutions, filling range 1μl~10ml. It has the advantages of high accuracy, high repeatability, high efficiency, high cleanliness, maintenance friendly and simple operation. It can be used manually and automatically.
Function and Feature
• High filling accuracy,the error is less than ±0.2%.
• Modular design,easy to expand,multiple pumps can be cascaded to form a multi-channel filling system.
• The man-machine interface directly input liquid volume parameters, easy to operate.
• High-precision servo motor drive,large torque,free-maintenance.
• Rotary pressure tube structure, high efficiency, small volume.
• The pump tube is low in loss, the continuous service life up to 700 hours, 12 hours attenuation rate<1%, simple maintenance.
• Suction back function,zero dripping,instant shutdown.
• High clean pipeline, easy to disassemble, easy to clean and sterilize, support CIP and SIP.
• It can be operated manually or used with an automatic production line.
Technical Parameters
Filling Range |
1μl~10ml |
Channel |
1(Single),up to 16 channels cascade are supported |
Filling Time |
0.01~99.99S |
Interval Time |
0.01~99.99S |
Filling Times |
0(Infinite loop)~9999 |
Tubing |
ID 0.5~9.6mm |
Display |
7 inch HD color touch screen |
External Control Signal |
Dry contact start signal, lack of bottle filling stop signal, feedback signal |
Communication |
RS485,MODBUS commnunication protocol |
Power |
Single channel<60W |
Power Supply |
AC100~240V,50/60Hz |
Dimension |
Reference size drawing |
Feature Comparision
Compared with Traditional Peristaltic Pump
|
MLP Peristaltic Pump
|
Traditional Peristaltic Pump
|
Filling Accuracy
|
high accuracy,high repeatability, filling10μl error<±0.2μl,100ul error<±1μl, 10ml error<±0.4%
|
not suitable for filling less than 100ul,200μl error<±3μl, 10ml error<±1%
|
Filling Efficiency
|
high efficiency,productivity up to 60 bottles/min
|
average efficiency,productivity<40 bottles/min
|
Attenuation
|
12 hours continuous operation,attenuation<1%,work one day without calibration
|
12 H continuous operation,attentuation>5%, need to calibrate every 2 H
|
Tube Life
|
continuous operation,service life>700 hours
|
fail after 70-100 hours of continuous operation
|
Pulsation
|
no pulse,small impact,no splash
|
large pulse, easy to splash
|
Compared with Plunger Pump
|
MLP Peristaltic Pump
|
Plunger Pump
|
Cleanliness
|
no dead volume,no effusion,material only contact with clean pipeline,no pollution
|
dead volume,effusion,material contact with the pump head cavity
|
Ease of Use
|
filling different material,only replace the pipeline to avoid cross contamination, low cost,simple maintenance
|
One pump can only fill one material, complicated maintenance
|
Universality
|
the material may contain particles such as magnetic beads
|
material must be clean and free of impurities
|
Dripping
|
retractable without dripping
|
easy to drip
|
Cavity Shedding
|
low extrusion strength,no mechanical friction,smooth cavity wall without shedding
|
continuous mechanical friction,the cavity is easy to fall off particles
|
Filling Accuracy and Efficiency
MLP500
Filling Range
|
Filling Time
|
Error Value
|
Accuracy Error
|
12 Hours Attenuation Rate
|
1μl
|
0.01~0.02
|
±0.1μL
|
<±10%
|
<1%
|
5μl
|
0.09~0.18s1
|
±0.1μL
|
<±2%
|
10μl
|
0.18~0.36
|
±0.1μL
|
<±1%
|
0.09~0.18
|
±0.2μL
|
<±2%
|
20μl
|
0.18~0.36
|
±0.2μL
|
<±0.5%
|
50μl
|
0.22~0.44
|
±0.5μL
|
<±1%
|
100μl
|
0.2~0.4
|
±1μL
|
<±1%
|
200μl
|
0.17~0.35
|
±2μL
|
<±1%
|
300μl
|
0.27~0.541
|
±2μL
|
<±0.6%
|
0.13~0.27
|
±4μL
|
<±1.3%
|
500μl
|
0.23~0.46
|
±4μL
|
<±0.8%
|
MLP3000
Filling Range
|
Filling Time
|
Error Value
|
Accuracy Error
|
12 Hours Attenuation Rate
|
1μl
|
0.01~0.02
|
±0.1μL
|
<±10%
|
<1%
|
5μl
|
0.04~0.12
|
±0.1μL
|
<±2%
|
10μl
|
0.09~0.23
|
±0.1μL
|
<±1%
|
20μl
|
0.18~0.46
|
±0.1μL
|
<±0.5%
|
0.1~0.24
|
±0.2μL
|
<±1%
|
50μl
|
0.25~0.6
|
±0.2μL
|
<±0.4%
|
0.08~0.2
|
±0.5μL
|
<±1%
|
100μl
|
0.16~0.4
|
±0.5μL
|
<±0.5%
|
0.13~0.3
|
1μL
|
<±1%
|
200μl
|
0.26~0.6
|
±1μL
|
<±0.5%
|
0.1~0.25
|
±2μL
|
<±1%
|
300μl
|
0.15~0.35
|
±2μL
|
<±0.6%
|
0.1~0.22
|
±4μL
|
<±1.6%
|
500μl
|
0.25~0.6
|
±2μL
|
<±0.4%
|
0.16~0.35
|
±4μL
|
<±0.8%
|
800μl
|
0.11~0.23
|
±8μL
|
<±1%
|
0.11~0.23
|
±8μL
|
<±1%
|
1ml
|
0.13~0.26
|
±8μL
|
<±0.8%
|
0.13~0.26
|
±8μL
|
<±0.8%
|
0.08~0.17
|
±13μL
|
<±1.3%
|
3ml
|
0.25~0.5
|
±13μL
|
<±0.6%
|
MLP10000
Filling Range
|
Filling Time
|
Error Value
|
Accuracy Error
|
12 Hours Attenuation Rate
|
100μl
|
0.15~0.3
|
±1μL
|
<±1%
|
|
200μ
|
0.25~0.5
|
±1μL
|
<±0.5%
|
300μl
|
0.4~0.8
|
±1μL
|
<±0.3%
|
500μl
|
0.2~0.4
|
±5μL
|
<±1%
|
1ml
|
0.17~0.35
|
±8μL
|
<±0.8%
|
0.1~0.2
|
±13μL
|
<±1.3%
|
3ml
|
0.3~0.6
|
±13μL
|
<±0.5%
|
0.2~0.4
|
±20μL
|
<±0.6%
|
5ml
|
0.35~0.7
|
±20μL
|
<±0.4%
|
2%
|
10ml
|
0.5~1
|
±40μL
|
<±0.4%
|
2%
|
Dimension UN(mm)
MLP500

MLP3000

MLP10000

bio-equip.cn
Baoding Lead Fluid Technology Co., Ltd. was established in October 1999. It is a high-tech joint-stock company, mainly engaged in peristaltic pumps,gear pumps,syringe pumps,OEM pumps research and development, production and sales; as well as related precision control of fluid, such as research and development, production and sales of stepper motor, drive and controller. We undertake all types of intelligent flow control system design, transformation, debugging and technical advice (including the overall program).
Peristaltic pumps are particularly useful in agriculture, biological technology, filtering, chemical industry, environment protection, the pharmaceutical industry, printing industry, water treatment industry,waste water treatment etc.
Our company is located in Baoding, Hebei Province,with convenient transportation of railroads and highways. We sincerely look forward to your visit and cooperation!
Website:http://leadfluidpump.com
Tel: 0086-312-3250677/877/977
Email:master@leadfluid.com / sale4@leadfluid.com
|