  • Product Detail
  • Company Profile

The Direct-Pure EDI lab water system produces Type II water from tap water directly. Resistivity of the product water is above 5 MΩ·cm at 25 ℃, which meets or exceeds Type II water quality as defined by ASTM, CAP, CLSI and ISO 3696 / BS 3997 and also complies with the Purified Water requirements from the European and U.S. Pharmacopoeia.
Features:
- The system incorporates the best in class Siemens EDI module to deliver high quality Type II water consistently at a relatively lower cost.
- Automatic water quality control programs ensure EDI product water meets or exceeds Type II water requirements.
- The fully automated control system contains pre-rinse, RO membrane fouling preventing flush, RO membrane life monitoring as well as many other programs for convenient operation.
- The pre-rinsing program starts automatically when a new RO membrane is installed so there is no need for users to set up additional cleaning procedures.
- Users can set water dispensing time for easy water collection.
- The system is equipped with a 30-liter PE tank (upgradable to a 60-liter) and a liquid level sensor and a tank vent filter with CO2 scavenger to prevent EDI water from carbon dioxide contamination. 
- RephiLe can provide full document support to meet users’ GMP, GLP, FDA and other certification requirements.
Main Applications:
- Type II Pure water can be used in many areas. Some of the typical applications are.
- Preparation of bio-reagents
- Preparation of culture media
- Preparation of solution for chemical analysis such as HPLC, ICP etc.
- For clinical analyzers
- Feed water for ultrapure water systems
- Medical device and equipment rinsing
- For serum and blood fractions
Specification:

 Feed Water Quality  
 *Tap water  Municipal water, TDS <1000 ppm (2000 μS/cm)
 Operating Temperature  5~35 ℃
 Feed Water Pressure  15 - 90 psi (1 - 6 bar)
 Water Production Rate  
 Type II water  5 or 10 L/h at 25 ℃
 Product Water Quality  
 Resistivity

 > 5 MΩ·cm (typically 10-16 MΩ·cm)

 TOC  < 30 ppb
 Particle (> 0.2 μm)  < 1/ml (with a 0.2 μm final filter)
 Microorganisms  < 0.1 cfu/ml (with a 0.2 μm final filter)
 Pyrogens (endotoxins)  < 0.001 Eu/ml (with a terminal ultrafiltration cartridge)
 Dimensions  
 Length x Depth x Height  30 cm × 51 cm × 48 cm
 Net Weight (Operating Weight)  22 - 25 kg (25 - 32 kg)

*If water hardness is less than 120 ppm, RephiLe suggests using RephiTEKT RW50211KT.

If water hardness is greater than 120 ppm, RephiLe suggests using an additional water softening device to protect the RO membrane in your system.

Ordering information:

 Description  Catalog No.
 Direct-Pure Water System, EDI 5  RD0E0050E
 Direct-Pure Water System, EDI 5 UV  RD0E005UE
 Direct-Pure Water System, EDI 10  RD0E0100E
 Direct-Pure Water System, EDI 10 UV  RD0E010UE
 Direct-Pure Water System, EDI 15  RD0E0150E
 Direct-Pure Water System, EDI 15 UV  RD0E015UE

Cartridge Set is required for the water systems to function. Please contact RephiLe for detailed information.

 

Contact us for more information

US: call toll-free 1-855-RephiLe (1-855-737-4453)

Other Countries: email info@rephile.com.  A local representative in your area will contact you.

bio-equip.cn
Request Infomation

* Name:
Title:
* Tel:
Fax:
* E-mail:
Postcode:
Institution/Company:
Address:
* Country:
Request infomation:
yes no
Request Quotation:
yes no
* Message:
Validated Code:
refresh
I agree to share my inquiry to the other matching suppliers.



Copyright(C) 2006-2025 Bio-Equip    E-mail:web@bio-equip.cn   沪ICP备06040519号